236 CHAPTER 20. ADVANCED DATA CALIBRATION<br />

1. method=‘fs’<br />

method: This method uses a fitting technique based on the Fouks-Schubert model. It is<br />

best used on an SSCD – see Section 20.3 for an example.<br />

There are three different implementations of this algorithm: an IDL version, a Fortran<br />

version and a C++ version. To use the IDL or Fortran version set the stabilize<br />

keyword /idl or /fortran.<br />

Some points to note about these different implementations:<br />

• The IDL version does not work well with irregularly gridded data, as can occur<br />

in the pixel history if a glitched pixel is recorded by the complex MASK. Consequently<br />

it is recommended to only use the simple MASK with the IDL version.<br />

See Section 2.3.4 for configuring the MASK.<br />

• The Fortran and C++ versions can handle irregularly gridded data. In this case<br />

both MASK settings are fine, though the complex MASK is recommended as it<br />

correctly indicates that glitched pixels should be ignored.<br />

routine called: corr transient fouks<br />

PDS side effects: Unstable pixels are flagged in .MASK and .CUBE is modified.<br />

reference: Coulais A. and Abergel A., 2000<br />

2. method=‘s90’<br />

method: Pixels which have not achieved 90% of their stable values are masked. This is<br />

the default stabilization method.<br />

routine called: corr transient s90<br />

PDS side effects: Unstable pixels are flagged in .MASK and .CUBE remains unmodified.<br />

reference: <strong><strong>ISO</strong>CAM</strong> Handbook, Chapter Data processing methods, Section Find the first<br />

stabilized pixel at each position without using a model.<br />

3. method=‘m90’<br />

method: This method applies a model to find the first stabilized pixel at each STATE.<br />

Pixels which have not achieved 90% of their stable values are masked.<br />

routine called: corr transient m90<br />

PDS side effects: Unstable pixels are flagged in .MASK and .CUBE remains unmodified.<br />

reference: <strong><strong>ISO</strong>CAM</strong> Handbook, Chapter Data processing methods, Section Find the first<br />

stabilized pixel at each position using a model.<br />

4. method=‘inv’<br />

method: This method uses a fitting technique called the IAS or INVERSION Model.<br />

This method is one of the more successful fitting methods.<br />

routine called: corr transient inv<br />

PDS side effects: Unstable pixels are flagged in .MASK and .CUBE is modified.<br />

reference: <strong><strong>ISO</strong>CAM</strong> Handbook, Chapter Data processing methods, Section Simplified<br />