304 CHAPTER 23. X CIA REFERENCE GUIDE<br />

• Process / Selected: Execute the user selected data processing. Warnings are displayed if<br />

some corrections have already been made to the current data set. Note that these warnings<br />

concern only the current x cia session. So before processing data, the user should have a<br />

look at the data history in order to check that some data corrections have not yet been<br />

processed (see above Data / Display History).<br />

• Tools / xv temp: Launch xv temp. See Section 14.4.7.<br />

• Tools / xv raster: Launch xv raster. See Section 14.4.11.<br />

• Tools / isocont (<strong><strong>ISO</strong>CAM</strong> axis): Launch isocont. See Section 14.6.2.<br />

• Tools / isocont (RA/DEC axis): As above, but the MOSAIC is displayed with EAST<br />

left and NORTH up.<br />

• Tools / x isocont: Launch x isocont. See Section 14.6.3.<br />

• Tools / hardcopy (gif): Create a print file of the isocont display (see above) in GIF<br />

format.<br />

• Tools / hardcopy (ps): Create a print file of the isocont display (see above) in postscript<br />

format.<br />

Some additional corrections can be performed. But second order corrections should not be<br />

used unless standard processing has been performed first to assess the quality of data.<br />

• Remove Dark Residuals: Dark pattern residuals can be removed using filtering in<br />

Fourier space. Note that this secondary order method updates <strong>ISO</strong>DATA.IMAGE, whereas<br />

first order dark correction updates <strong>ISO</strong>DATA.CUBE.<br />

• Smooth Undefined Values: It may happen that <strong>ISO</strong>DATA.RASTER contains undefined<br />

values (if there is no stabilized value for example). Undefined values can be replaced<br />

with smoothed values, as determined from neighboring pixels.<br />

• FOV Distortion: The FOV distortion effect can be corrected: the trapezoidal shape<br />

of pixels is then taken into account in order to re-sample the image. Distortion can be<br />

assumed to be linear around the central pixel (linear method) or can be calculated from<br />

ground calibration measurements (calibration method).<br />

• Spectral Deconvolution: The measured spectrum can be deconvolved of the spectral<br />

resolution of the CVF filter. Note that the spectral sensitivity is corrected by selecting the<br />

Jy/pixel units instead of ADU/s/pixel.