20.11.2012 Views

AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...

AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...

AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

AA Business Unit<br />

<strong>Service</strong> Information<br />

Author: Bernhard Conrads No: <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

Tel.: +49-7551-81-3720 Pages: 68<br />

e-mail: bernhard_conrads@eur.perkin-elmer.com Date: 09. May 2000<br />

This document contains proprietry information and may only be used by the recipient for the prescribed purposes and<br />

may be neither reproduced in any form nor the document itself or its contents divulged to third parties without our<br />

expressed prior written permission. All rights reserved.<br />

<strong>AS</strong>-<strong>93plus</strong> <strong>Service</strong> <strong>Manual</strong><br />

Problem<br />

N/A<br />

Purpose<br />

The autosampler model <strong>AS</strong>-91 stand-alone P/N B050-9090 has been discontinued and<br />

replaced with the new model <strong>AS</strong>-<strong>93plus</strong>. This document shall provide the field with<br />

service information.<br />

Note This revision "B" of the <strong>AS</strong>-<strong>93plus</strong> <strong>Service</strong> <strong>Manual</strong> contains significant<br />

changes of the rinse kit which have been introduced to provide a better support<br />

of ICP/ICP-MS applications plus a few minor corrections. The previous revision<br />

"A" of this document is obsolete and should be discarded.<br />

Material Required<br />

N/A<br />

General<br />

Procedure<br />

N/A<br />

Note Refer to SDB <strong>AS</strong>9N009A.<strong>STB</strong> for electrical schematics as the PC Boards<br />

respectively the electrical schematics are identical with the autosampler<br />

<strong>AS</strong>-90plus.<br />

Note Refer to SDB <strong>AS</strong>9N011A.SPB containing the illustrated spare parts list.<br />

<strong>AS</strong>-<strong>93plus</strong> <strong>Service</strong> <strong>Manual</strong> No: <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Table of Content<br />

Contents page<br />

Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1<br />

Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1<br />

Material Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1<br />

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1<br />

Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1<br />

Table of Content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2<br />

Introduction and System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4<br />

Matched to the customers need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4<br />

Key benefits: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Fast, accurate random access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Large, flexible sample capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Standard rack configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Corrosion-resistant sampling components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Automatic rinsing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5<br />

Sample racks and sample trays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6<br />

The following sample tray groups can be selected: . . . . . . . . . . . . . . . . . . . . . . . . . 6<br />

Further sample racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8<br />

Tray definition files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8<br />

Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8<br />

Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9<br />

Unpacking and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10<br />

Carton recycling instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10<br />

Checklist of the <strong>AS</strong>-<strong>93plus</strong> Accessory Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11<br />

Checklist of the Rinsing Kit B314-0236 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12<br />

Assembling sample racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13<br />

Positioning the autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14<br />

Electrical connectors and switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15<br />

The DIP switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16<br />

To set the DIP switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16<br />

Example for setting an IEEE address: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16<br />

Selecting the IEEE-488 interface on the <strong>AS</strong>-<strong>93plus</strong>: . . . . . . . . . . . . . . . . . . .17<br />

Selecting the RS-232C interface on the <strong>AS</strong>-<strong>93plus</strong>: . . . . . . . . . . . . . . . . . . . .17<br />

Connecting the autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17<br />

To fit the sampling probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18<br />

Adjusting the sampling probe height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20<br />

Connecting the sample tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21<br />

To connect the sample tube to the sampling probe . . . . . . . . . . . . . . . . . . . . . . . . . 21<br />

To connect the sample tube to the nebulizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22<br />

Normal connection procedure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22<br />

Alternate connection procedure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23<br />

Installing the sample racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24<br />

Installing PerkinElmer sample racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24<br />

Sample rack locations for the sample tray groups . . . . . . . . . . . . . . . . . . . . . . . . . . 25<br />

Installing Gilson sample racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28<br />

Setting up the rinsing port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30<br />

Connecting the tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30<br />

Pump tubes and retaining hooks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30<br />

Tubing configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31<br />

2 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Configuration 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32<br />

Configuration 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34<br />

Configuration 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36<br />

Fitting pump tubes to the rinse pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38<br />

Installing the rinsing port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39<br />

Function test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40<br />

<strong>AS</strong>-<strong>93plus</strong> Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43<br />

Installation of the <strong>AS</strong>-93 utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43<br />

Start the <strong>AS</strong>-93 utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43<br />

Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44<br />

Tray Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45<br />

How to create a tray definition file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47<br />

Rinse Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49<br />

Firmware Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50<br />

Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51<br />

Alignment of the <strong>AS</strong>-<strong>93plus</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52<br />

Alignment using TERMINAL.EXE and ADJ93.TRM . . . . . . . . . . . . . . . . . . . . . . . . . . . .52<br />

Kinds of positional misalignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56<br />

Correction of a parallel misalignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56<br />

Correction of a X/Y-offset misalignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58<br />

Checking the Event Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61<br />

Internal Cyclic tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61<br />

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63<br />

Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64<br />

Direct commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64<br />

Identification Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64<br />

Reset Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64<br />

Tray commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64<br />

Pump Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65<br />

X-, Y-, Z- Drive commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65<br />

Event Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66<br />

<strong>AS</strong>-<strong>93plus</strong> Error Code Reference Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 3 of 68


Introduction and System Description<br />

Matched to the customers need<br />

The PerkinElmer <strong>AS</strong>-<strong>93plus</strong> autosampler is a computer-controlled, multi-purpose<br />

sampling system for flame AA, ICP-OES and ICP-MS. The <strong>AS</strong>-<strong>93plus</strong> automates<br />

standard and sample introduction for instrument calibration and sample analysis,<br />

extending the spectrometer’s capabilities to those of a fully automated analytical<br />

workstation. The <strong>AS</strong>-<strong>93plus</strong> is computer-controlled using PerkinElmer AA WinLab,<br />

ICP WinLab or ELAN ® software.<br />

The autosampler is operated from the keyboard and screen of the spectrometer system;<br />

the autosampler itself does not have any operating controls.<br />

The autosampler does not incorporate any device for active sample aspiration but merely<br />

moves its sampling probe into the individual containers with the calibration and sample<br />

solutions. For solution aspiration the sampling probe must be connected via a sample<br />

tube to an active module. An active module can, for example, be the nebulizer of an A<strong>AS</strong><br />

burner, the pump of a flow injection system, or the pump of an ICP nebulizer assembly.<br />

Tower<br />

Sampling probe<br />

Rinsing<br />

port<br />

Plug-in<br />

power supply<br />

. .<br />

.<br />

Rinse<br />

pump<br />

Figure 1 Schematic of the Autosampler<br />

Autosampler arm<br />

Sample table<br />

Sample tube<br />

Removable<br />

sample racks<br />

4 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Key benefits:<br />

• fast, accurate random access<br />

• large, flexible sample capacity<br />

• standard rack configurations<br />

• corrosion-resistant sampling components<br />

• automatic rinsing<br />

Fast, accurate random access<br />

An advanced drive system moves the sampling arm of the <strong>AS</strong>-<strong>93plus</strong> in the X and Y<br />

coordinates simultaneously, minimizing changeover time between samples. High<br />

precision digital stepper motors move the <strong>AS</strong>-<strong>93plus</strong> sampling arm to its programmed<br />

X-Y-Z coordinates quickly and accurately. Random access programming gives the user<br />

exceptional flexibility in locating different samples and calibration solutions.<br />

Large, flexible sample capacity<br />

Sample capacity with the <strong>AS</strong>-<strong>93plus</strong> is up to 200 samples per sample rack. Sample racks<br />

are easily interchanged, reducing analysis set-up times and giving the user the flexibility<br />

to use differently sized sample vessels matched to his analytical needs.<br />

Standard rack configurations<br />

In addition to the variety of sample racks offered by PerkinElmer, the <strong>AS</strong>-<strong>93plus</strong> is also<br />

compatible with commonly used laboratory sample rack configurations from other<br />

manufacturers, such as Scienceware or Gilson.<br />

Corrosion-resistant sampling components<br />

The sample probe assembly, including the sample transport capillary tubing, is made<br />

entirely of acid- and solvent-resistant PTFE.<br />

Automatic rinsing<br />

The autosampler includes a flow-through rinsing port so that the sampling probe is always<br />

rinsed with fresh rinse liquid, significantly reducing the risk of carryover. This is especially<br />

advantageous when samples with very different element concentrations have to be<br />

analyzed. Rinse liquid is pumped in a stream through the rinsing port by a peristaltic pump<br />

that is built into the autosampler. The pump operates during the rinse phase of the<br />

measurement cycle (i.e. every time that the sampling probe enters the rinsing port) and is<br />

idle during other phases of the cycle. You can select or change for how long the rinse<br />

pump is switched on via the Rinse download utility.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 5 of 68


Sample racks and sample trays<br />

The autosampler offers exceptional sample handling capacity. Various, easily removable,<br />

sample racks allow a variety of differently sized sample containers to be used. Three<br />

PerkinElmer sample racks can be installed on the autosampler. Within the WinLab<br />

software a sample tray is defined as a group of three racks.<br />

The following sample tray groups can be selected:<br />

Tray group E<br />

Rinsing port<br />

location<br />

B314-0621 B314-0617 (2x)<br />

This group (option) comprises the following sample racks:<br />

• Rack B314-0621 (rinsing-port rack) with rinsing port location (location 0) and 20<br />

locations for 50 mL solution containers (calibration and/or test sample<br />

solutions).<br />

• Rack B314-0617 (2x) each with 90 locations for 6 mL and 8 mL solution<br />

containers (test sample solutions).<br />

6 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Tray group F<br />

This group (provided as standard) comprises the following sample racks:<br />

• Rack B300-1647 (rinsing-port rack) with rinsing port location (location 0),<br />

8 locations for 50mL solution containers (calibration and/or test sample<br />

solutions), and 29 locations for 15 mL solution containers (test sample<br />

solutions).<br />

• Rack B314-0618 (2x) each with 60 locations for 15mL solution containers (test<br />

sample solutions).<br />

Tray group G<br />

Rinsing port<br />

location<br />

Rinsing port<br />

location<br />

B300-1647 B314-0618 (2x)<br />

B314-0621 (3x)<br />

This group (option) comprises the following sample racks:<br />

• Rack B314-0621 (3x) with 21 locations each for 50mL solution containers<br />

(calibration and/or test sample solutions).<br />

Location 0 in the first rack is for the rinsing port.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 7 of 68


Further sample racks<br />

In addition to the above sample racks, you can also use racks offered by Gilson. To install<br />

the Gilson racks on the autosampler you require an adapter plate (option). Four Gilson<br />

racks can be installed on the autosampler. Please note that the Gilson racks cannot be<br />

obtained from PerkinElmer.<br />

Further sample racks can be offered for special applications. Please refer to the current<br />

price lists for information.<br />

Tray definition files<br />

A tray definition file provides the system with information about the group of sample racks<br />

installed on the autosampler. The system requires this information to move the sampling<br />

probe to the correct location for each container in the rack. You select the required sample<br />

tray group in the spectrometer’s operating software when you set up the autosampler.<br />

The spectrometer operating software already contains tray definition files for the<br />

PerkinElmer sample tray groups E, F, and G.<br />

Tray definition files are also included with the autosampler for each of the sample tray<br />

groups that can be selected.<br />

Further information on tray definition files is provided on page 47.<br />

Quick Reference<br />

A slot is provided underneath the autosampler into which you can insert the<br />

Quick Reference. We recommend that you keep the Quick Reference with the<br />

autosampler at all times since it contains maintenance information and a list of<br />

replacement parts.<br />

8 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Technical data<br />

System design Compact computer-controlled system for automatic solution<br />

withdrawal for atomic spectrometric applications.<br />

System control Control is via the same software that is used to control the<br />

spectrometer system.<br />

Chemical resistance The autosampler surfaces are resistant to dilute acids and<br />

alkalis, and to a lesser extent to strong acids and alkalis and<br />

organic solvents.<br />

Autosampler arm<br />

movement<br />

Vertical:145mm<br />

Horizontal:X-axis (left–right)302mm<br />

Y-axis (front–back)222mm<br />

Resolution:0.1 mm ± 4%<br />

Speed:typically 3s from one location to the next.<br />

RS232C and IEEE-488.<br />

Communication<br />

interfaces<br />

Power requirements 40V; provided by the plug-in power supply.<br />

Electrical protection As defined in EN 61010-1-1993 (IEC 1010-1):<br />

Insulation: Class I<br />

Insulation category (overvoltage category): II<br />

Pollution degree: 2<br />

Safety and EMC As defined in EN 61010-1-1993 (IEC 1010-1: 1990 + A1 +<br />

standards A2) and<br />

EN 61010-2-061:1996.<br />

Electromagnetic compatibility: EN 61326-1:1997.<br />

Environmental Ambient temperature: +15 °C to +35 °C,<br />

requirements 20–80% relative humidity, non-condensing.<br />

Dimensions 420mm wide x 370mm high x 333mm deep.<br />

Mass (weight) approx. 4kg<br />

Specifications of the plug-in power supply<br />

Intended use<br />

Electrical<br />

requirements<br />

Plugs<br />

Output<br />

Safety standards<br />

EMC standards<br />

Power supply for information technology and similar<br />

equipment.<br />

100–240 VAC, 47–63Hz<br />

The plug-in power supply is provided with 4 interchangeable<br />

plugs compliant with European, UK, USA, and Australian<br />

standards.<br />

40VDC, 700mA<br />

As defined in EN 60950.<br />

EN 55011, EN 55014, and EN 55022/B;<br />

EN 6100-3-2, -3, EN 6100-4-2, -3, -4, -5, -6, -11.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 9 of 68


Unpacking and inspection<br />

The autosampler is shipped in a single packing carton.<br />

After you receive the autosampler, unpack the components carefully:<br />

1. Open the carton and remove the bags containing the sample containers.<br />

2. Remove the top foam piece from the carton.<br />

3. Take out the small parts and then remove the foam piece.<br />

4. The autosampler is now accessible.<br />

Carefully lift it out and place it on the bench.<br />

Do not lift it by the autosampler arm or the tower.<br />

5. Remove packing material from the autosampler.<br />

6. Examine the components for any signs of damage during shipment.<br />

In the event of damage, file an immediate claim with the authorized carrier,<br />

and inform your local PerkinElmer office or representative.<br />

A list of parts provided is given on next page.<br />

Carton recycling instructions<br />

Caution: Risk of damage to the instrument<br />

• Never lift the autosampler by the autosampler arm or tower.<br />

Tower<br />

Autosampler arm<br />

The carton can be folded down for easy storage. You can store the carton for future use<br />

(to transport the instrument), recycle it by returning it to PerkinElmer, or break it down<br />

completely and give it to your local recycling center.<br />

10 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Checklist of the <strong>AS</strong>-<strong>93plus</strong> Accessory Kit<br />

Table 1 Checklist of the <strong>AS</strong>-<strong>93plus</strong> accessory kit<br />

Qty. Item Part No.<br />

1 Plug-in power supply 0999-8283<br />

1 Communication cable (autosampler to spectrometer) B314-4034<br />

1 Sample rack B300-1647<br />

2 Sample racks B314-0618<br />

1 Sampling probe B300-0055<br />

1 Sampling tube (0.6 mm i.d.) with screw fittings B300-0158<br />

1 Sample tube (1 mm i.d.) with screw fittings B019-1060<br />

1 Adapter F B019-3873<br />

2 Polyethylene bottles, 2000 mL B010-4344<br />

1 Cleaning wire (40cm) B050-5962<br />

1 Rinsing kit (refer to table 2 for details) B314-0236<br />

1 Pack of 50 sample containers, 15 mL,<br />

polypropylene with screw caps<br />

1 Pack of 50 sample containers, 50 mL,<br />

polypropylene with screw caps<br />

B050-0662<br />

B050-1397<br />

1 User’s Guide 0993-5278<br />

1 Quick Reference 0993-5279<br />

1 <strong>AS</strong>-93 Utility Disk B314-0758<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 11 of 68


Checklist of the Rinsing Kit B314-0236<br />

Table 2 The rinsing kit B314-0236 includes the following items:<br />

Qty. Item Part No.<br />

1 Rinsing port, pre-assembled B314-0622<br />

1 Pump tube (pack of six), collars red-red, 1.14 mm i.d. B314-0730<br />

1 Pump tube (pack of six), purple-white, 2.79 mm i.d. B314-0721<br />

1 m Tygon® drain tube 6.4 x 1.6 B050-9650<br />

1 m PVC tube B004-8139<br />

0.1 m PVC tube B004-8139<br />

0.1 m Silicone tube B001-8283<br />

1 Rinse liquid feed tube with weight B019-1059<br />

1 Connector, reducing 3.2 mm to 1.7 mm B019-9233<br />

2 Connector, 4.5 mm to 4.5 mm B314-0715<br />

1 Adapter A B019-3342<br />

1 Adapter C B019-6850<br />

1 Adapter E B019-6857<br />

2 Adapter M B050-7919<br />

1 Silicone Oil, Bottle of 5 mL B314-0341<br />

12 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Assembling sample racks<br />

For easy shipment, sample racks are often folded flat. Follow the procedure below to<br />

assemble a sample rack.<br />

Edge tab<br />

End plate<br />

Lugs<br />

Slot tabs<br />

Figure 2 Assembling a Sample Rack<br />

Certain sample racks have numbered tiers (more information: see ‘Sample tray group F’,<br />

page 26 and see ‘Sample tray group G’, page27 . When you assemble a rack, make sure<br />

that the set of numbers that you require faces upward on the top tier.<br />

1. Place the base on the bench so that the lugs are underneath.<br />

2. Bend up the two end plates.<br />

End plate<br />

Top tier<br />

Middle tier<br />

3. Take the middle tier and snap the edge tabs into the lower set of slots<br />

in the end plates.<br />

4. Bend up the slot tabs on the sides of the base and snap the edge tabs<br />

of the middle tier into the slots.<br />

5. Assemble the top tier into the respective slots in the same manner.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 13 of 68<br />

Base


Positioning the autosampler<br />

The autosampler requires the same environmental conditions as the spectrometer.<br />

Please refer to the guide provided with your spectrometer for details.<br />

When used alone, the autosampler is connected directly to the nebulizer’s sample tube.<br />

Since the length of the sample tube has an influence on the aspiration rate of the<br />

nebulizer, you should position the autosampler as close to the spectrometer as possible.<br />

The best position is on a suitable table or trolley (cart) in front of the spectrometer’s<br />

atomizer compartment. With some spectrometers it is possible to place the autosampler<br />

on a sample tray directly in front of the atomizer compartment.<br />

If you are using a diluter system, such as the AutoPrep-50, you can place this next to the<br />

autosampler to keep the sample tubes as short as possible.<br />

Autosampler<br />

Spectrometer<br />

Nebulizer<br />

Sample tube<br />

Figure 3 Suggested Arrangement for the System (1)<br />

14 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Diluter<br />

(option)<br />

Autosampler<br />

Figure 4 Suggested Arrangement for the System (2)<br />

Electrical connectors and switches<br />

The connectors and switches are located on the underside of the autosampler.<br />

Table 3 Electrical connectors and switches<br />

Item Designation Function<br />

Spectrometer<br />

Nebulizer<br />

Sample tube<br />

Autosampler sample tube<br />

1 2 3 4 5<br />

1 Power Power input socket; for 40V supply from plug-in power supply.<br />

2 Event Potential-free relay contact to activate an external system.<br />

3 Com 2 RS232C communication port.<br />

4 –– DIP switches (see next page for more details).<br />

5 Com 1 IEEE-488 communication port.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 15 of 68


The DIP switches<br />

You use the DIP switches to select the required communication parameters such as the<br />

interface and, where applicable, the instrument address for the IEEE interface.<br />

For the <strong>AS</strong>-<strong>93plus</strong> the DIP switches have the following functions and values:<br />

To set the DIP switches<br />

ON<br />

1 2 3 4 5 6 7 8<br />

The DIP switches<br />

Switch Function Value<br />

ON OFF<br />

(up) (down)<br />

1 1 0<br />

2 Switches 1–4 set the instrument address for the 2 0<br />

3 IEEE-488 communication interface.<br />

4 0<br />

4 8 0<br />

5 Switches 5, 6 and 7 select service programs for use OFF<br />

6 by a PerkinElmer engineer. In normal operation OFF<br />

7 they must be set to OFF (down).<br />

OFF<br />

8 Switch 8 selects between the RS-232C and the<br />

IEEE-488 interfaces.<br />

IEEE RS-232<br />

Caution: Risk of damage to the instrument<br />

Electronic components may be damaged if you change the DIP<br />

switch settings when the instrument is switched on.<br />

• Switch off the instrument before you change the DIP switch<br />

settings.<br />

Altered settings on the DIP switches become effective the next time you switch the<br />

instrument on.<br />

Example for setting an IEEE address:<br />

You add the values of the DIP switches to obtain the address. To select an IEEE address<br />

of 14, for example, set the DIP switches 1–4 as shown below:<br />

ON<br />

1 2 3 4 5 6 7 8<br />

Address = 14<br />

(= 0+2+4+8)<br />

16 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


The usual addresses preselected for the autosampler in the spectrometer operating software are:<br />

AA WinLab 14, 3, or 7<br />

ICP WinLab 14<br />

ICP-MS 7<br />

Selecting the IEEE-488 interface on the <strong>AS</strong>-<strong>93plus</strong>:<br />

1. Set the DIP switches as follows:<br />

Selecting the RS-232C interface on the <strong>AS</strong>-<strong>93plus</strong>:<br />

1. Set all DIP switches to OFF.<br />

Switches 5, 6, 7, and 8 must be set to OFF. Switches 1 through 4 can be in<br />

any arbitrary position, but we recommend that you set them to OFF.<br />

Connecting the autosampler<br />

1–4 Set to the address selected for the<br />

autosampler (e.g. 14).<br />

5–7 Set to OFF.<br />

8 Set to ON.<br />

This section describes how you connect the autosampler to your spectrometer system.<br />

The procedure for connecting the spectrometer and the computer are described in the<br />

spectrometer guide.<br />

Warning: Electrical Hazard<br />

• To prevent potential injury to yourself and damage to the instrument, switch<br />

OFF all instruments in the system and disconnect them from the line power<br />

supply before you perform the procedures described below.<br />

1. Switch OFF the spectrometer and other components in the system and<br />

disconnect them from the line power supply.<br />

2. Place the autosampler in front of the spectrometer’s atomizer compartment<br />

see ‘Positioning the autosampler’, page14.<br />

3. Select IEEE-488 communication and the address on the DIP switches, see<br />

‘The DIP switches’, page16.<br />

4. Connect the communication cable (B314-4034), provided, between the Com1<br />

port on the underside of the autosampler and the IEEE-488 communication port<br />

at the spectrometer (refer to the Spectrometer Guide for details). The cables<br />

are provided with stacked connectors.<br />

5. Check that the plug fitted to the plug-in power supply is suitable for the<br />

electrical outlets in your laboratory.<br />

If it is not:<br />

1. Slide off the plug fitted to the power supply.<br />

2. Slide in the correct plug, making sure that it clicks fully into place.<br />

Four interchangeable plugs are provided, compliant with European, UK, USA, and<br />

Australian standards.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 17 of 68<br />

W2.3


Communication<br />

cable<br />

(B314-4034)<br />

Plug-in<br />

power supply<br />

Figure 5 Electrical Connections: <strong>AS</strong>-<strong>93plus</strong> Autosampler to Spectrometer with IEEE<br />

Communication<br />

6. Connect the system to the line power supply:<br />

1. Connect the autosampler to the line power supply using the plugin<br />

power supply provided. The plug-in power supply uses a widerange<br />

input of 100–240V, 47–63Hz.<br />

2. Connect the spectrometer and other components in the system to<br />

the line power supply.<br />

Note To avoid problems of interference caused by earth loops, always connect<br />

the spectrometer, computer, printer and other components to the same phase of<br />

the line power supply. The most convenient method is to use a multisocket<br />

distributor.<br />

To fit the sampling probe<br />

IEEE-488<br />

communication port<br />

Power<br />

Spectrometer<br />

Autosampler<br />

Com 1<br />

7. Switch on all components of the system, following the routine in the<br />

Spectrometer Guide.<br />

After you have completed the installation and electrical connections of the autosampler,<br />

fit the sampling probe as follows:<br />

18 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Caution: Risk of damage to the autosampler components<br />

• When you fit the sampling probe, take care not to press<br />

forcefully in any direction on the autosampler arm since this can<br />

damage the arm.<br />

• The sampling probe is a complete assembly. Do not attempt to<br />

take the probe apart since this can damage it; reassembly is not<br />

possible.<br />

1. If not already done, carefully insert the sampling probe guide from above into<br />

the sampling probe holder on the autosampler arm and push it down fully.<br />

2. Carefully insert the sampling probe from above into the probe guide.<br />

Push the sampling probe down until about 20 mm protrude from the bottom<br />

end of the probe guide.<br />

Use the O-ring on the probe to prevent it from slipping down.<br />

After you have fitted the sampling probe, adjust the height as described in the next<br />

section.<br />

Note The standard sampling probe is a PTFE-lined epoxy polymer tube. This<br />

polymer is resistant to the acids, alkalis and organic solvents normally used in<br />

atomic spectroscopy. If your samples contain an organic solvent that attacks this<br />

polymer, a PTFE-lined stainless steel sampling probe is offered as an option<br />

(P/N B300-0152).<br />

Tower<br />

Sampling probe guide<br />

Autosampler arm<br />

.<br />

Holder<br />

approx.<br />

20 mm<br />

Figure 6 Fitting the Sampling Probe<br />

Sample tube<br />

Screw connector<br />

O-ring<br />

Sampling probe<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 19 of 68


Adjusting the sampling probe height<br />

Note The procedure described below assumes that you are using a<br />

spectrometer operated via the AA WinLab software. If you are using an ICP or<br />

ICP-MS instrument, please refer to the information on operating an autosampler<br />

in the online help for the respective instrument.<br />

1. Install the rinsing-port sample rack on the autosampler, see ‘Installing the<br />

sample racks’, page24.<br />

2. Place the rinsing port in the rinse location (location zero) in the sample rack<br />

see ‘Setting up the rinsing port’, page30.<br />

3. On the Analyze page, click on Select Location.<br />

Select Go to wash and then click on OK.<br />

4. The sampling probe moves to above the rinsing port and is then lowered.<br />

5. Carefully push the sampling probe down until it bottoms in the rinsing port.<br />

Then pull the sampling probe up by about 5mm. Use the O-ring on the probe<br />

to prevent it from slipping down.<br />

The sampling probe is now at the correct height.<br />

6. To remove the sample rack, first click on Load Tray to raise the sampling probe,<br />

then remove the rack.<br />

This completes installation of the sampling probe. Connect the sample tube between the<br />

sampling probe and the spectrometer’s sample input system as described on the following<br />

pages.<br />

Sampling probe<br />

Rinsing port<br />

5 mm<br />

Figure 7 Adjusting the Sampling Probe Height<br />

20 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

.


Connecting the sample tube<br />

The sample uptake rate is governed by the internal diameter of the sample tube. Two<br />

sample tubes with different internal diameters are provided with the autosampler. Each<br />

sample tube has a screw connector on one end for connection to the sampling probe and<br />

a screw connector on the other end for connection to the nebulizer via an adapter.<br />

You use the sample tube with 0.6mm internal diameter (B300-0158) for routine samples.<br />

You use the sample tube with 1mm internal diameter (B019-1060) if you are operating<br />

the burner system with a High Performance Nebulizer; this sample tube allows the<br />

nebulizer to attain its maximum performance.<br />

To connect the sample tube to the sampling probe<br />

Tower cover<br />

.<br />

Depending on how you position the autosampler with respect to the rest of the system,<br />

you can attach the sample tube to either the front of the autosampler arm or the rear of<br />

the tower cover. A spring clip with two slots is provided to hold the sample tube; the<br />

narrower slot holds the 0.6 mm tube, while the wider slot holds the 1mm tube.<br />

1. Screw the connector on the sample tube finger-tight into the fitting on<br />

the sampling probe.<br />

2. Make a smooth loop in the sample tube and then attach it to either the<br />

front of the autosampler arm or the rear of the tower cover:<br />

1. Insert the spring clip into the appropriate socket.<br />

2. Thread the sample tube gently around the hook in the correct slot.<br />

Note Do not kink or pinch the sample tube.<br />

Spring clip<br />

with two slots<br />

Note The loop must be large enough so that the sampling probe holder can<br />

move unhindered backward and forward along the autosampler arm.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 21 of 68


To connect the sample tube to the nebulizer<br />

The following procedure describes how you connect the sample tube to the nebulizer of<br />

an A<strong>AS</strong> burner system. If you are using an ICP or ICP-MS spectrometer system, please<br />

refer to the guide provided with the system for information on how to connect the sample<br />

tube to the nebulizer-spray chamber assembly.<br />

Normal connection procedure:<br />

Note Do not kink or pinch the sample tube.<br />

1. Take the free end of sample tube from the autosampler to the<br />

nebulizer.<br />

2. Connect the sample tube to the nebulizer:<br />

1. Shorten the nebulizer’s existing sample tube (capillary) so that the<br />

length is about 50 mm.<br />

2. Connect adapter F (B019-3873, provided) to the shortened sample<br />

tube.<br />

3. Screw the connector on the autosampler sample tube into adapter<br />

F.<br />

3. Make certain that the sampling probe can move freely to all locations<br />

in the sample racks.<br />

Shortened nebulizer sample tube<br />

Autosampler<br />

sample tube with<br />

screw connector<br />

Adapter F<br />

Nebulizer<br />

22 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Alternate connection procedure:<br />

Adapter F contains a Pt/Ir capillary. If you are aspirating very aggressive samples<br />

(containing hydrofluoric acid for example) or if the samples should not contact metal, you<br />

can connect the sample tube directly to the nebulizer without using the adapter. The<br />

procedure is described below.<br />

Note Do not kink or pinch the sample tube. When you cut it, make a clean cut<br />

with a scalpel – do not use scissors.<br />

1. Take the free end of sample tube from the autosampler to the<br />

nebulizer.<br />

Cut off the screw connector from the end of the sample tube.<br />

If the sample tube is longer than necessary, shorten it to the required<br />

length, but make certain that it is long enough so that the sampling<br />

probe can move freely to all locations in the sample racks.<br />

Note An 0.6mm i.d. sample tube with only one connector is offered as an option<br />

(P/N B300-0157). With this sample tube it is not necessary to cut off the screw<br />

connector.<br />

2. Connect the free end of the sample tube to the nebulizer:<br />

1. Standard nebulizers:<br />

Connect the sample tube to the nebulizer inlet,<br />

using a cleaning wire to aid installation (refer to<br />

the Spectrometer Guide for details).<br />

- or -<br />

2. GemTip nebulizers:<br />

Insert the sample tube into the PTFE tube located<br />

in the gland nut of the nebulizer (refer to the<br />

Spectrometer Guide for details).<br />

Gland nut<br />

Insert sample tube<br />

into PTFE tube<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 23 of 68


Installing the sample racks<br />

Installing PerkinElmer sample racks<br />

For easy shipment, the sample racks are often folded flat. The procedure for assembling<br />

the racks is described in ’Assembling sample racks on page 13’.<br />

To install the racks:<br />

Caution: Risk of damage to the instrument<br />

If the sample racks installed on the autosampler are not the correct<br />

ones for the sample tray selected in the WinLab software, the<br />

sampling probe may strike against a rack rather than enter a<br />

sample location.<br />

• Always make sure that the correct racks are installed.<br />

Note The procedures described below assumes that you are using a<br />

spectrometer operated via the AA WinLab software. If you are using an ICP or<br />

ICP-MS instrument, please refer to the information on operating an autosampler<br />

in the online help for the respective instrument.<br />

1. Start AA WinLab (or the operating software specific to the<br />

spectrometer).<br />

2. Select the sample racks that make up the sample tray group you<br />

intend to use more information: page 25.<br />

3. In the WinLab software, select the required sample tray group:<br />

1. Open the correct method.<br />

2. Display the Method Editor window.<br />

3. On the Calibration - Autosampler page, select the correct group of<br />

sample racks under Tray Type (E, F, or G).<br />

4. Load the sample racks with the calibration and sample solutions.<br />

Carefully note the location of each solution in the racks see ‘Sample<br />

rack locations for the sample tray groups’, page25.<br />

5. If required, drive the autosampler arm to the load tray position:<br />

1. Display the Automated Analysis window.<br />

2. On the Analyze page, click on Load Tray.<br />

If the autosampler arm is not currently in the load tray position, the arm rises<br />

and is moved to the origin at the rear of the sample table. In this position the<br />

sampling probe does not hinder installation or removal of the sample racks.<br />

24 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


6. Install the sample racks on the autosampler:<br />

1. Place the rinsing-port rack on the left-hand side of the sample table<br />

such that the rinsing port is at the rear left (as seen from the front<br />

of the autosampler). Make sure that the lugs on the bottom of the<br />

sample rack fit into the slots in the sample table.<br />

2. Place the other two sample racks in the correct sequence on the<br />

sample table.<br />

Note Sample racks can be placed on the sample table in two directions.<br />

Make certain that you place each sample rack in the proper direction so that the<br />

solutions are worked down in the correct sequence.<br />

7. To remove sample racks, first click on Load Tray to raise the sampling<br />

probe, then remove the racks.<br />

Sample rack locations for the sample tray groups<br />

The rack locations for the various sample tray groups are depicted on the following pages.<br />

The autosampler works down the sample locations in numerical sequence.<br />

Sample tray group E<br />

Rinsing<br />

port<br />

0<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

Rack B314-0621:21 locations (30 mm diam.) for 50mL containers,<br />

one location (0) can be used for the rinsing port.<br />

Rack B314-0617:90 locations (13 mm square) for 6mL or 8 mL sample containers.<br />

Tray definition file for this group of racks: <strong>AS</strong>93E200.tra<br />

21<br />

22<br />

23<br />

24<br />

25<br />

26<br />

27<br />

28<br />

29<br />

30<br />

31<br />

32<br />

33<br />

34<br />

35<br />

36<br />

37<br />

38<br />

39<br />

40<br />

41<br />

42<br />

43<br />

44<br />

45<br />

46<br />

47<br />

48<br />

49<br />

50<br />

51<br />

52<br />

53<br />

54<br />

55<br />

56<br />

57<br />

58<br />

59<br />

60<br />

61<br />

62<br />

63<br />

64<br />

65<br />

66<br />

67<br />

68<br />

69<br />

70<br />

71<br />

72<br />

73<br />

74<br />

75<br />

76<br />

77<br />

78<br />

79<br />

80<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 25 of 68<br />

81<br />

82<br />

83<br />

84<br />

96<br />

97<br />

98<br />

99<br />

85 100<br />

86 101<br />

87 102<br />

88 103<br />

89 104<br />

90 105<br />

91 106<br />

92 107<br />

93 108<br />

94 109<br />

95 110<br />

Rack B314-0621 Rack B314-0617 (2x)<br />

111 126141<br />

156 171 186<br />

112 127142<br />

157 172 187<br />

113 128143<br />

158 173 188<br />

114 129144<br />

159 174 189<br />

115 130145<br />

160 175 190<br />

116 131146<br />

161 176 191<br />

117 132147<br />

162 177 192<br />

118 133148<br />

163 178 193<br />

119 134149<br />

164 179 194<br />

120 135150<br />

165 180 195<br />

121 136151<br />

166 181 196<br />

122 137152<br />

167 182 197<br />

123 138153<br />

168 183 198<br />

124 139154<br />

169 184 199<br />

125 140155<br />

170 185 200


Note The proprietary sample racks are not marked. We suggest that you mark<br />

the racks to denote the first location in the rack for easy orientation when installing<br />

the racks on the autosampler.<br />

Sample tray group F<br />

Rinsing<br />

port<br />

0<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

21<br />

22<br />

23<br />

24<br />

25<br />

Rack B300-1647:9 locations (30mm diam.) for 50mL containers, one location (0) can be<br />

used for the rinsing port; 29 locations (16 mm diam.) for 15mL sample containers.<br />

Rack B314-0618:60 locations (16mm diam.) for 15 mL sample containers.<br />

Tray definition file for this group of racks: <strong>AS</strong>93F157.tra<br />

26<br />

27<br />

28<br />

29<br />

30<br />

31<br />

32<br />

33<br />

34<br />

35<br />

36<br />

37<br />

38<br />

39<br />

40<br />

41<br />

42<br />

43<br />

44<br />

45<br />

46<br />

47<br />

48<br />

49<br />

50<br />

51<br />

52<br />

53<br />

54<br />

55<br />

56<br />

57<br />

58<br />

59<br />

60<br />

61<br />

Rack B300-1647 Rack B314-0618 (2x)<br />

62<br />

63<br />

64<br />

65<br />

66<br />

67<br />

68<br />

69<br />

70<br />

71<br />

72<br />

73<br />

Note The top tier of rack B314-0618 is numbered on both sides; one side is<br />

numbered 38–97 and the other side is numbered 98–157. When you assemble a<br />

rack you can turn the tier over to obtain the set of numbers that you require.<br />

26 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

74<br />

75<br />

76<br />

77<br />

78<br />

79<br />

80<br />

81<br />

82<br />

83<br />

84<br />

85<br />

86<br />

87<br />

88<br />

89<br />

90<br />

91<br />

92<br />

93<br />

94<br />

95<br />

96<br />

97<br />

98 110 122 134 146<br />

99<br />

111<br />

123<br />

135<br />

147<br />

100 112 124 136 148<br />

101 113 125 137 149<br />

102 114 126 138 150<br />

103 115 127 139 151<br />

104 116 128 140 152<br />

105 117 129 141 153<br />

106 118 130 142 154<br />

107 119 131 143 155<br />

108 120 132 144 156<br />

109 121 133 145 157


Sample tray group G<br />

Rinsing<br />

port<br />

0<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

Rack B314-0621:21 locations (30 mm diam.) for 50mL containers,<br />

one location (0) can be used for the rinsing port.<br />

Tray definition file for this group of racks: <strong>AS</strong>93G62.tra<br />

21<br />

22<br />

23<br />

24<br />

25<br />

26<br />

27<br />

28<br />

29<br />

30<br />

31<br />

32<br />

33<br />

34<br />

Note The tiers of rack B314-0621 are numbered. When you assemble a rack<br />

you can put the tier with the required set of numbers at the top.<br />

One tier is numbered 0–20; use this as the top tier in the rinsing-port rack. Make<br />

sure that you assemble the rack with the rinsing-port location (0) above the socket<br />

for the rinsing port in the base.<br />

The second tier is numbered on both sides; one side is numbered 21–41 and the<br />

other side is numbered 42–62. You can turn the tier over to obtain the set of<br />

numbers that you require.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 27 of 68<br />

35<br />

36<br />

37<br />

38<br />

39<br />

40<br />

41<br />

Rack B314-0621 (3x)<br />

42<br />

43<br />

44<br />

45<br />

46<br />

47<br />

48<br />

49<br />

50<br />

51<br />

52<br />

53<br />

54<br />

55<br />

56<br />

57<br />

58<br />

59<br />

60<br />

61<br />

62


Installing Gilson sample racks<br />

In addition to the PerkinElmer sample racks, you can also use racks offered by Gilson. Up<br />

to four Gilson racks can be installed on the autosampler. Within the WinLab software a<br />

sample tray is defined as a group of four racks. To install the Gilson racks on the<br />

autosampler you require an adapter plate (option; P/N B314-0680). Please note that the<br />

Gilson racks cannot be obtained from PerkinElmer.<br />

To install the racks:<br />

1. If necessary, download the required tray definition files more<br />

information: page 45.<br />

2. Prepare the autosampler as described under To install the racks: on<br />

page 24 (see step 1 through step 5).<br />

3. Place the adapter plate (option; P/N B314-0680) on the sample table.<br />

Pins on the underside of the adapter plate engage in slots on the<br />

sample table.<br />

4. Place rack type 24 (29mm diam.) on the left-hand side of the sample<br />

table.<br />

Note You must always place rack type 24 on the left-hand side. Other racks<br />

cannot be used in this position.<br />

5. Place the other three racks on the sample table.<br />

Figure 8 Placing Gilson Racks on the Adapter Plate (autosampler table not shown)<br />

28 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Sample tray groups for Gilson racks<br />

1 rack type 24<br />

(29 mm diam.)<br />

3 racks type 21<br />

(13.5mm diam.)<br />

Tray definition file for<br />

this group of racks:<br />

<strong>AS</strong>93gilson193.tra<br />

1 rack type 24<br />

(29 mm diam.)<br />

3 racks type 22/23<br />

(18.5mm diam.)<br />

Tray definition file for<br />

this group of racks:<br />

<strong>AS</strong>93gilson145.tra<br />

4 racks type 24<br />

(29 mm diam.)<br />

Tray definition file for<br />

this group of racks:<br />

<strong>AS</strong>93gilson055.tra<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 29 of 68


Setting up the rinsing port<br />

Connecting the tubing<br />

Pump tubes and retaining hooks<br />

The pump tubes have three collars, giving two sections that can be tensioned around the<br />

pump rotor. If one section of a tube becomes deformed, and there is no other damage,<br />

you can use the other section.<br />

Pump tubes of different internal diameter are available so that you can vary the flow rate<br />

of the rinse liquid through the rinsing port. The pump tubes have colored collars for easy<br />

identification.<br />

Pump tube<br />

Color<br />

(collars)<br />

The retaining hooks on the rinse pump have an inside channel for the rinse feed pump<br />

tube and an outside channel for the drain pump tube (see Figure 9). There are slots for the<br />

pump tubes on the top and bottom of each retaining hook; the slots on the top are for tubes<br />

with narrower diameter, while the slots on the bottom are for tubes with wider diameter.<br />

Figure 9 Rinse Pump<br />

i.d. (mm)<br />

B314-0730 red-red 1.14 3.5–5<br />

B314-0721 violet-white 2.79 11–14<br />

Rear channel<br />

Front channel<br />

Pump rotor<br />

Section 1 Section 2<br />

Top view<br />

Collars Pump tube<br />

Retaining hook<br />

Nominal flow<br />

rate (mL/min)<br />

Side view<br />

Retaining hook<br />

Retaining hook<br />

Pump<br />

rotor<br />

30 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Tubing configurations<br />

You can set up several different tubing configurations to meet the given requirements.<br />

• Configuration 1. The rinse pump is used to pump rinse liquid into the rinsing port<br />

and to pump the liquid back out into the waste bottle. A narrow-bore pump tube<br />

(B314-0730; red-red) is used for the rinse feed to provide a slow flow rate of the<br />

rinse liquid. This is the preferred configuration for ICP-OES and ICP-MS<br />

applications.<br />

• Configuration 2. The rinse pump is used to pump rinse liquid into the rinsing port<br />

and to pump the liquid back out into the waste bottle. A wide-bore pump tube<br />

(B314-0721; violet-white) is used for the rinse feed to provide a fast flow rate of<br />

the rinse liquid. This is the preferred configuration for flame A<strong>AS</strong> applications.<br />

• Configuration 3. The rinse pump is used to pump rinse liquid into the rinsing port<br />

only. The drain tube is connected to the rinsing port directly and the liquid drains<br />

by gravity into the waste bottle. A wide-bore pump tube (B314-0721; violetwhite)<br />

is used for the rinse feed to provide a fast flow rate of the rinse liquid.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 31 of 68


Configuration 1<br />

Rinsing<br />

port<br />

Rinse liquid<br />

feed line<br />

Bottle with<br />

rinse liquid<br />

Figure 10 Configuration 1 – Schematic of Tubing Connections<br />

Items required for configuration 1<br />

1 2 3<br />

4<br />

8<br />

5<br />

= Color-coded collars on pump tubes<br />

Drain line<br />

Waste bottle<br />

Rinse<br />

pump<br />

Item Description Part No.<br />

1 Adapter E, 6.3mm o.d. barbed fitting B019-6857<br />

2 Adapter M, 4mm o.d. barbed fitting B050-7919<br />

3 Wide-bore pump tube, 2.79mm i.d., violet-white B314-0721<br />

4 Connector, reducing 3.2mm to 1.7mm B019-9233<br />

5 Narrow-bore pump tube, 1.14mm i.d., red-red B314-0730<br />

6 Connector, 4.5mm to 4.5mm o.d. barbed fittings B314-0715<br />

7 Adapter A, 1.8mm o.d. barbed fitting B019-3342<br />

8 Rinse liquid feed tube B019-1059<br />

9 PVC drain tube B004-8139<br />

32 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

9<br />

6 7


Connecting the tubes for configuration 1<br />

1. Fit the rinse feed pump tube (red-red) (5) and the drain pump tube<br />

(violet-white) (3) to the rinse pump as described in Fitting pump tubes<br />

to the rinse pump on page 38.<br />

Note Always fit a wide-bore drain pump tube (violet-white).<br />

2. Connect up the rinse feed line:<br />

1. Push the wider barbed fitting of the reducing<br />

connector (4) into the PVC tube prefitted to the inlet<br />

fitting of the rinsing port.<br />

2. Push the end of the rinse feed pump tube ( 5) coming<br />

from the upper retaining hook onto the narrower<br />

barbed fitting of the reducing connector (4).<br />

3. Push the barbed fitting of adapter A ( 7) into the rinse<br />

feed pump tube coming from the lower retaining<br />

hook.<br />

4. Screw the connector on the rinse liquid feed tube (8)<br />

into adapter A.<br />

Tube connector<br />

Adapter<br />

3. Connect up the drain line:<br />

1. Screw adapter M (2) into adapter E (1).<br />

2. Push the barbed fitting of adapter E into the silicone tube prefitted<br />

to the drain nozzle on the rinsing port.<br />

3. Push the end of the drain tube (3) coming from the lower retaining<br />

hook onto the barbed fitting of adapter M.<br />

4. Push the tube connector (6) into the end of the drain pump tube<br />

coming from the upper retaining hook.<br />

5. Push one end of the PVC drain tube (9) onto the connector.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 33 of 68


Configuration 2<br />

Rinsing<br />

port<br />

Rinse liquid<br />

feed line<br />

Bottle with<br />

rinse liquid<br />

Figure 11 Configuration 2 – Schematic of Tubing Connections<br />

Items required for configuration 2<br />

1 2 3<br />

4<br />

8<br />

5<br />

= Color-coded collars on pump tubes<br />

Drain line<br />

Waste bottle<br />

Rinse<br />

pump<br />

Item Description Part No.<br />

1 Adapter E, 6.3mm o.d. barbed fitting B019-6857<br />

2 Adapter M, 4mm o.d. barbed fitting B050-7919<br />

3 Wide-bore pump tube, 2.79mm i.d., violet-white B314-0721<br />

4 Connector, 4.5mm to 4.5mm o.d. barbed fittings B314-0715<br />

5 Wide-bore pump tube, 2.79mm i.d., violet-white B314-0721<br />

6 Connector, 4.5mm to 4.5mm o.d. barbed fittings B314-0715<br />

7 Adapter C, 4mm o.d. barbed fitting B019-3342<br />

8 Rinse liquid feed tube B019-1059<br />

9 PVC drain tube B004-8139<br />

34 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

9<br />

6 7


Connecting the tubes for configuration 2<br />

1. Fit the rinse feed pump tube (violet-white) (5) and the drain pump tube<br />

(violet-white) (3) to the rinse pump as described in Fitting pump tubes<br />

to the rinse pump on page 38.<br />

Note Always fit a wide-bore drain pump tube (violet-white).<br />

2. Connect up the rinse feed line:<br />

1. Push a barbed fitting of the tube connector (4) into<br />

the PVC tube prefitted to the inlet fitting of the<br />

rinsing port.<br />

2. Push the end of the rinse feed pump tube ( 5) coming<br />

from the upper retaining hook onto the other barbed<br />

fitting of the tube connector (4).<br />

3. Push the barbed fitting of adapter C (7) into the rinse<br />

feed pump tube coming from the lower retaining<br />

hook.<br />

4. Screw the connector on the rinse liquid feed tube (8)<br />

into adapter C.<br />

Tube connector<br />

Adapter<br />

3. Connect up the drain line:<br />

1. Screw adapter M (2) into adapter E (1).<br />

2. Push the barbed fitting of adapter E into the silicone tube prefitted<br />

to the drain nozzle on the rinsing port.<br />

3. Push the end of the drain tube (3) coming from the lower retaining<br />

hook onto the barbed fitting of adapter M.<br />

4. Push the tube connector (6) into the end of the drain pump tube<br />

coming from the upper retaining hook.<br />

5. Push one end of the PVC drain tube (9) onto the connector.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 35 of 68


Configuration 3<br />

Rinsing<br />

port<br />

Drain line<br />

Waste bottle<br />

5<br />

Figure 12 Configuration 3 – Schematic of Tubing Connections<br />

Items required for configuration 3<br />

= Color-coded collars on pump tube<br />

Rinse liquid<br />

feed line<br />

Bottle with<br />

rinse liquid<br />

Rinse<br />

pump<br />

Item Description Part No.<br />

1 Connector, 4.5mm to 4.5mm o.d. barbed fittings B314-0715<br />

2 Wide-bore pump tube, 2.79mm i.d., violet-white B314-0721<br />

3 Adapter C, 4mm o.d. barbed fitting B019-3342<br />

4 Rinse liquid feed tube B019-1059<br />

5 Tygon ® drain tube B050-9650<br />

36 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

1<br />

2<br />

4<br />

3


Connecting the tubes for configuration 3<br />

1. Fit the rinse feed pump tube (violet-white) (2) to the rinse pump as<br />

described in Fitting pump tubes to the rinse pump on page 38.<br />

2. Connect up the rinse feed line:<br />

1. Push a barbed fitting of the tube connector (1) into<br />

the PVC tube prefitted to the inlet fitting of the<br />

rinsing port.<br />

2. Push the end of the rinse feed pump tube ( 2) coming<br />

from the upper retaining hook onto the other barbed<br />

fitting of the tube connector (1).<br />

3. Push the barbed fitting of adapter C (3) into the rinse<br />

feed pump tube coming from the lower retaining<br />

hook.<br />

4. Screw the connector on the rinse liquid feed tube (4)<br />

into adapter C.<br />

3. Connect up the drain line:<br />

1. Remove the prefitted piece of silicone tube from the drain nozzle<br />

on the rinsing port.<br />

2. Push one end of the Tygon drain tube (5) onto the drain nozzle.<br />

3. To ensure smooth draining, make sure that the drain tube hangs<br />

straight down without loops or kinks.<br />

Lug<br />

Drain Nozzle<br />

Rinse liquid<br />

inlet fitting (adapter M, prefitted)<br />

Figure 13 Rinsing Port (prefitted tubes not shown)<br />

Tube connector<br />

Adapter<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 37 of 68


Fitting pump tubes to the rinse pump<br />

1. Swing the cam lever fully<br />

clockwise to release the<br />

pressure block.<br />

2. Pull the pressure block off<br />

the pivot.<br />

3. Fit the rinse feed pump<br />

tube:<br />

1. Hook a collar into the<br />

inside channel of a<br />

retaining hook, using<br />

the correct slot for the<br />

diameter of the tube.<br />

2. Gently tension the pump<br />

tube around the pump<br />

rotor and hook the<br />

second collar into the<br />

inside channel of the<br />

other retaining hook.<br />

Do not twist or kink the<br />

pump tube.<br />

4. Fit the drain pump tube:<br />

1. Repeat substeps 1 and<br />

2, above, to fit the drain<br />

pump tube to the<br />

outside channel of the<br />

retaining hooks.<br />

Collars<br />

5. Place a drop of silicone oil on the pump tubes in contact with the pump<br />

rollers to reduce wear.<br />

6. Slide the pressure block back onto the pivot.<br />

Retaining hook<br />

Retaining hook<br />

Cam lever<br />

Pressure<br />

block<br />

7. Swing the cam lever anti-clockwise to press the pressure block<br />

against the pump tubes.<br />

The final position of the cam lever depends on the pressure that is<br />

required on the pump tubes for satisfactory operation (the “3 o’clock”<br />

position is usually about right). Check the flow of liquid through the<br />

rinsing port when the pump is operating. If the flow is intermittent or<br />

inadequate, increase the pressure on the pump tubes with the cam<br />

lever. The lifetime of the pump tubes will be shortened if the pressure<br />

is too high.<br />

Note: Release the pressure on the pump tubes during longer pauses in operation (e.g.<br />

overnight) to prolong the lifetime of the pump tubes.<br />

Make sure that the inner diameter of the drain pump tube is the same as, or larger than,<br />

the rinse feed pump tube. If the drain pump tube has a smaller inner diameter, rinse<br />

liquid will back up in the rinsing port and ultimately overflow into the autosampler,<br />

possibly causing damage.<br />

38 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

Pivot<br />

Pump tube<br />

Pump<br />

rotor


Installing the rinsing port<br />

Lug passes<br />

into slot<br />

(not visible)<br />

Apertures<br />

Figure 14 Rinsing Port Installed in the Rack (tubes not shown)<br />

1. Orientate the rinsing port such that the tubes will pass through the<br />

apertures in the rack.<br />

2. Lower the rinsing port into the rack and guide the lug on the side of the<br />

rinsing port into the slot in the rack.<br />

3. Install the sample rack on the autosampler and position the<br />

autosampler close to the edge of the bench so that the drain tube<br />

hangs straight down.<br />

4. Place the waste bottle on a shelf or on the floor underneath the<br />

instrument in a position where it will not be kicked or knocked over.<br />

Make sure that the waste bottle is lower than the rinsing port.<br />

5. Place the free end of the drain tube into the waste bottle. If necessary,<br />

shorten the drain tube; to ensure smooth draining it must hang straight<br />

down without loops into the waste bottle and the end of the tube must<br />

never be immersed in liquid in the waste bottle. Make sure that the<br />

drain tube cannot slip out of the bottle.<br />

6. Place the rinse liquid bottle in a convenient position adjacent to the<br />

instrument where it cannot be knocked over. The bottle should be<br />

lower than the rinsing port to prevent siphoning.<br />

7. Place the free end of the rinse liquid feed tube (with the weight) into<br />

the rinse liquid bottle. Make sure that the tube cannot slip out of the<br />

bottle.<br />

Note You can install and remove the rinsing port without having to disconnect<br />

the tubes. When you remove the rinsing port from the rack, empty the liquid in the<br />

port into a beaker to prevent possible spillages in the laboratory.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 39 of 68


Function test<br />

The sense of the following test is to determine the mechanical accuracy of the X- and Y-<br />

sample probe drives. The follow description is based on the use of AAWinLab 3.0 under<br />

Windows 95 with an AAnalyst-100:<br />

1. Ensure, that the PC, the spectrometer and the <strong>AS</strong>-<strong>93plus</strong> autosampler<br />

are correctly connected and powered on.<br />

2. Start Windows 95 as usual and wait until the Windows initialization<br />

has been completed.<br />

3. Click on "Start" in the Windows task bar, then click on "Programs".<br />

4. In the program group "AAnalyst 100" double click the icon<br />

"Reconfigure". The "Reconfigure" starts and will show it’s first window.<br />

5. Click on the radio button of the used spectrometer (e.g. AAnalyst 100).<br />

"Reconfigure" will show the next window as shown below:<br />

Figure 15 Installation Option Window<br />

6. Click on the radio buttons indicating the hardware configuration of<br />

your system. Figure 13 shows the smallest possible configuration,<br />

e.g. AA-100 and <strong>AS</strong>-<strong>93plus</strong> only.)<br />

Note If the Installation Option windows does not show the radio button<br />

<strong>AS</strong>-<strong>93plus</strong>, select the radio button <strong>AS</strong>-91. The <strong>AS</strong>-<strong>93plus</strong> is operational reverse<br />

compatible to the <strong>AS</strong>-91.<br />

40 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


7. Click on "Next" to finish the reconfiguration procedure.<br />

8. Double click the Icon "AAWinLab Analyst" to start the reconfigured<br />

spectrometer operating software. After a while the AAWinLab’s initial<br />

screen will be shown with it’s menu line and toolbar.<br />

9. Select "File" - "New" to create a method to test the function and<br />

precision of the X- and Y-drive of the <strong>AS</strong>-<strong>93plus</strong>.<br />

10.Select an element (e.g. Cu) from the "Create New Method" window<br />

and click "OK".<br />

11.Type in the method description. Then select the tab "Calib." and click<br />

on the sub-tab "Autosampler".<br />

Figure 16 Autosampler Window<br />

12.Select the "Tray Type" (rack group). and select "F" as this is the<br />

standard rack configuration of the <strong>AS</strong>-<strong>93plus</strong>.<br />

Note Figure 14 shows the 3 types of predefined tray types (=rack groups) using<br />

Perkin Elmer racks. The character E,F and G of the tray type designation<br />

identifies the rack group while the following numbers points out the number of the<br />

maximal sample locations of the tray type. Refer to page 25 for details about predefined<br />

"tray types" = rack groups.<br />

If you are using sample racks from other suppliers you have to copy appropriate<br />

"tray files" that are named according the 8.3 character file name convention into<br />

the AA Winlab/Elan directory. "Ready-to-use" tray -files are provided with the <strong>AS</strong>-<br />

93 utility disk or must be prepared as described on page 47.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 41 of 68


Note In case that you are operating the <strong>AS</strong>-<strong>93plus</strong> with an AA Winlab software<br />

revision older than rev. 3.0 (AAnalyst 100/300) respectively rev. 3.9 (AAnalyst<br />

600/700/800) or with a 16 bit ICP WinLab software revision, you have to<br />

download the appropriate tray files from the <strong>AS</strong>-93 utility disk as described in the<br />

next chapter.<br />

13.Minimize the "method editor" for later use.<br />

14.Click on the "Auto" button of the tool bar.<br />

AAWinLab will show the "Automated Analysis" window.<br />

15.Select the sub-window "Analyze" from the "Automated Analysis"<br />

window.<br />

16.Click on the "Select Location" button from the "Analyze" window.<br />

AAWinLab will show the "Autosampler" window.<br />

Figure 17 Autosampler Window<br />

17.Select "Go to location 0" and click on "OK" to start the <strong>AS</strong>-<strong>93plus</strong>.<br />

The sampling probe will move upwards, the tower and the sampling<br />

probe holder will then move to the reference position, i.e. the left rear<br />

corner. In the next moment the tower and the sampling probe holder<br />

will move to location "0" and the sampling probe will move downwards.<br />

Finally the peristaltic pump will start for 30 seconds.<br />

18.Use the "Go to location xx" function to move the sampling probe to all<br />

corner locations of each sample rack and check the mechanical<br />

precision. For the standard tray type F the corner locations are as<br />

follows:<br />

1. Left rack: location 0, 5, 26, 37<br />

2. Middle rack: location 38, 49, 86, 97<br />

3. Right rack loaction 98, 109, 146, 157.<br />

Note The pump is only running at location 0.<br />

19.Check the precision of all sample locations mentioned above.<br />

A tolerance of ± 1 mm is within specification<br />

Refer to chapter "Alignment of the <strong>AS</strong>-93" starting on page 52 if the<br />

tolerance exceeds 1mm.<br />

42 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


<strong>AS</strong>-<strong>93plus</strong> Utilities<br />

The <strong>AS</strong>-93 utilities allows to change certain operating parameters of the autosampler.<br />

Installation of the <strong>AS</strong>-93 utilities<br />

It is recommended to install the <strong>AS</strong>-93 utilities on the computer that is used to control the<br />

spectrometer system.<br />

1. Insert the <strong>AS</strong>-93 utilities disk into the drive.<br />

Usually the Utilities program will start automatically.<br />

(If the Setup program does not start automatically:<br />

press the Windows Start-button and click on Run in the Windows start menu.<br />

In the Run window click on Browse. In the Browse window, select your drive,<br />

select Setup.exe, and click on Open. In the Run window click on OK.)<br />

2. Follow the instructions on the screen to install the <strong>AS</strong>-93 utilities program.<br />

3. You might create a short cut to the <strong>AS</strong>-93 utilities program.<br />

Start the <strong>AS</strong>-93 utilities<br />

1. Shut down the flame, graphite furnace or ICP torch and then exit the<br />

spectrometer’s operating software.<br />

2. The autosampler must remain connected to the electrical supply.<br />

3. Start the <strong>AS</strong>-93 utilities:<br />

If you did create an icon during installation, double click the icon.<br />

4. If an icon is not available, start from Windows program menu:<br />

Start - Programs - <strong>AS</strong>-93 Utilities.<br />

5. Click on OK and the <strong>AS</strong>-93 Utilities window will appear on the screen.<br />

The <strong>AS</strong>-93 Utilities window consist of 5 tabs respectively sub-windows:<br />

• Communication<br />

• Tray Download<br />

• Rinse Download<br />

• Firmware Download<br />

• Info<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 43 of 68


Communication<br />

The Communication dialog allows to set up the communication type and parameters for<br />

the communication of the Utilities program with the autosampler. The communication<br />

type/parameters have to match with the dip-switch settings of the autosampler. If the<br />

communication parameters are not correct, you will not be able to download any of the<br />

utilities.<br />

To change the communication parameters:<br />

1. In the Communication dialog:<br />

1. Select the required Communication Interface:<br />

Click on IEEE or RS232.<br />

2. If you select IEEE, enter the IEEE Address for the autosampler.<br />

–or–<br />

If you select RS232, select the communication port (COM port) for<br />

the autosampler.<br />

The settings are implemented as soon as you change to a different page or close the<br />

window.<br />

Figure 18 <strong>AS</strong>-93 Utilities: Communication Dialog<br />

44 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Tray Download<br />

A tray definition file provides the system with information about the group of sample racks<br />

installed on the autosampler. This information includes the coordinates of each location<br />

in the sample rack. The system requires this information to move the sampling probe to<br />

the correct location for each sample container in the rack. Refer to page 48 for further<br />

information about location coordinates. The spectrometer operating software already<br />

provides tray definition files for the Perkin Elmer sample tray type E, F and G. Tray<br />

definition files for each of the sample tray types that can be selected are also provided<br />

with the autosampler. The most recent software versions of the AA Winlab software such<br />

as Rev.3.00 or higher (used to control the AAnalyst 100/300), Rev.3.90 or higher (used<br />

to control the AAnalyst 600/700/800), the 32 bit ICP WinLab software and the ICP-MS<br />

ELAN software make direct use of tray definition files. In this case the tray definition file<br />

must be located in the WinLab/Elan directory, i.e. a tray download by means of the <strong>AS</strong>-93<br />

Utility software is NOT required. A tray download as described below is only required in<br />

the case, that the <strong>AS</strong>-<strong>93plus</strong> is controlled by software versions not supporting tray<br />

definition files, e.g. 16 bit ICP WinLab software or AA WinLab software Rev. 2.61.<br />

Note Tray definition files for the use with AA WinLab software and ICP-MS<br />

ELAN software are available from the <strong>AS</strong>-93 Utility disk P/N B314-0758<br />

Setting up tray definition files<br />

If you change the types of sample racks for use (e.g. you want to use Gilson racks instead<br />

of PerkinElmer racks), you must set up the tray definition files for those sample racks. In<br />

the spectrometer operating software you can select tray type E, F, or G. The Tray<br />

Download dialog allows to define which tray definition files are assigned to E, F and to G.<br />

Note Any change of the rack configuration assigned to a tray type requires a<br />

download of the correct tray definition files.<br />

Performing a tray download<br />

Performing a tray download<br />

1. Make sure that the communication parameters have been set<br />

correctly.<br />

1. In the Tray File dialog click on Open for Tray E.<br />

2. In the Open window, in Look in ..., select the directory containing<br />

the tray definition files, click on the file (group of sample racks) that<br />

you wish to assign to Tray E and then click on Open.<br />

3. Repeat this procedure for Tray F and Tray G, if required.<br />

Note: You can assign one, two, or three tray definition files. It is not<br />

necessary to always assign three.<br />

2. In the Tray File dialog, click on Download to implement the tray<br />

definition files in the spectrometer’s operating software.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 45 of 68


Figure 19 <strong>AS</strong>-93 Utilities: Tray Download Dialog<br />

46 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


How to create a tray definition file<br />

With an text editor, such like Notepad, edit or write a new file as shown below:<br />

Figure 20 Tray Definition File of Tray Type " E"<br />

Figure 20 shows the tray definition file for the PerkinElmer tray type E. In this example, to<br />

save place, the text is cut out between the sample location 18 and 47.<br />

Note The MaxLocation parameter (maximum location number in tray) must be<br />

equal with the highest position number of the positions listing.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 47 of 68


The sample positions are defined as follows:<br />

Each location of the tray definition file is defined by X-, Y-, and Z-coordinates of the<br />

sample rack. The X-coordinates determine the left-right movement of the sampling probe,<br />

the Y-coordinates determine the front-rear movement, and the Z-coordinates determine<br />

the vertical movement.<br />

The origin position is identical with the initialization (home) position of the sampling probe<br />

at the rear of the sample table. The sampling probe returns to this position when the Load<br />

Tray command has been performed.<br />

Figure 21 Principle of sample locations and coordinates<br />

In the various tray definition files, the X- and Y-coordinates for each location in the sample<br />

racks are the distances (in mm) from the origin to the center of the location.<br />

For example, in the PerkinElmer sample racks the rinsing port (location 0) has the<br />

coordinates X = 9 mm and Y = 13 mm.<br />

You can inspect the location coordinates of a tray definition file by means of a text editor<br />

such as Microsoft Notepad (included with MS Windows).<br />

The ability to create customized tray definition files allows the use of customized sample<br />

racks or sample racks supplied from other vendors such as Gilson.<br />

48 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Rinse Download<br />

This utility enables you to change the rinse parameters for the autosampler.<br />

You can change:<br />

• The pump speed; this permits you to increase or decrease the flow of rinse<br />

liquid through the rinsing port.<br />

• the rinse time; you can increase or decrease the rinse time.<br />

• the rinse location; you can select a different location for the rinse liquid.<br />

The default parameters are shown in the entry boxes in the Rinse window. To display the<br />

values currently programmed for the autosampler, click on Show Parameters.<br />

Figure 22 <strong>AS</strong>-93 Utilities: Rinse Download Dialog<br />

To change rinse parameters:<br />

1. Make sure that the communication parameters have been set correctly.<br />

2. Make any changes that you require to the Pump speed and/or Rinse time.<br />

3. If you wish to use a different location for the rinse solution, enter the X- and Ycoordinates<br />

for the selected location (see ‘Location coordinates for sample<br />

racks’ in the User’s Guide).<br />

4. When you have completed all changes, click on Download to implement the<br />

changes in the spectrometer’s operating software.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 49 of 68


Firmware Download<br />

This utility enables you to download new firmware for the autosampler. At shipment, the<br />

<strong>AS</strong>-93 utilities disk provided with the autosampler contains the same firmware as installed<br />

in the autosampler. If PerkinElmer changes the firmware at any time, new firmware will be<br />

made available on the Lotus Notes AA <strong>Service</strong> database for download.<br />

To download firmware:<br />

1. Make sure that the communication parameters have been set<br />

correctly.<br />

2. Click on Browse.<br />

3. In the Open window:, In Look in ..., select the directory containing the<br />

firmware files.<br />

4. Select the file and click on Open.<br />

5. In the Firmware Download window, click on Download to install the<br />

new firmware in the autosampler.<br />

Figure 23 <strong>AS</strong>-93 Utilities: Firmware Download Dialog<br />

50 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Information<br />

The Information window displays information about the currently-installed version of the<br />

firmware for the autosampler.<br />

To display version information:<br />

1. Make sure that the communication parameters have been set correctly.<br />

2. Click on the Info tab.<br />

3. Click on Show Info to display the current version of the firmware.<br />

Figure 24 <strong>AS</strong>-93 Utilities: Information Windows<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 51 of 68


Alignment of the <strong>AS</strong>-<strong>93plus</strong><br />

This chapter describes the procedure to perform a mechanical alignment of the <strong>AS</strong>-<strong>93plus</strong><br />

mechanics.<br />

The <strong>AS</strong>-<strong>93plus</strong> is misaligned if the sampling probe does not hit the center of the sample<br />

cups. A misalignment might occur during transportation or wrong handling. A re-alignment<br />

might also be required after a repair of electronic components of the sampler.<br />

In general, adjustable are only X- and Y-position. The limits of the up and down movement<br />

of the autosampler arm are not adjustable. However, the height of the sampling probe<br />

need to be adjusted. This procedure is described on page 20.<br />

The <strong>AS</strong>-<strong>93plus</strong> offers an easy way to correct a misalignment by entering offset steps for<br />

the X- and Y-drive.<br />

The offset steps can be entered by using the utilities Terminal.exe and Adj93.trm or by<br />

entering of "direct commands" in the <strong>AS</strong>-93 device control module (= base module).<br />

Alignment using TERMINAL.EXE and ADJ93.TRM<br />

The "Terminal.exe" program with the setup file "Adj93.trm" allows the control of the<br />

<strong>AS</strong><strong>93plus</strong> by using "direct commands" via the RS 232 communication. The same direct<br />

commands are used by all actual application software. Only with the Terminal.exe it is<br />

possible to change the offset steps for alignment.<br />

Required parts:<br />

• PC with Windows 95 or NT<br />

• RS 232 communication cable (9 pole) B050-7701 with the following<br />

pin assignment:<br />

15 pin D-Sub male connector¸15 pin D-Sub female connector<br />

1 ¸ 1<br />

2 ¸ 2<br />

3 ¸ 3<br />

4 ¸ 4<br />

5 ¸ 5<br />

6 ¸ 6<br />

7 ¸ 7<br />

8 ¸ 8<br />

9 ¸ 9<br />

10 ¸ 10<br />

11 ¸ 11<br />

12 ¸ 12<br />

13 ¸ 13<br />

14 ¸ 14<br />

15 ¸ 15<br />

• Floppy disk with the program Terminal.exe and Adj93.trm<br />

(available from AA <strong>Service</strong> BBS Lotus Notes DB))<br />

Perform the follow procedure:<br />

52 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


1. Carefully turn over the autosampler to access the connector plate at<br />

the bottom (refer to illustration on page 15) and set all dip switches to<br />

OFF.<br />

2. Connect the cable of the 40 V plug-in power supply (see item 1 of<br />

illustration on page 15).<br />

3. Connect the RS232 cable from the autosampler communication port<br />

2 to the PC communication port 1 as shown below.<br />

Figure 25 RS232 Connection for <strong>Service</strong> Purpose<br />

4. Turn the autosampler back to the normal operating position. Make<br />

sure that the tower movement isn’t blocked. Place the sample racks<br />

(sample tray type F, standard accessory) on the autosampler. Insert<br />

the rinsing port (refer to page 39). Don't insert the sampling probe yet.<br />

5. Plug the "plug-in power supply" of the autosampler into the mains<br />

receptacle and switch on the PC. Allow the autosampler to complete<br />

the initialization (approx. 15 sec.).<br />

6. Click on the "Start" button of the PC and select the Windows Explorer.<br />

7. Insert the floppy disk with the "Terminal.exe" program.<br />

8. In the Explorer window, select drive A and double click the file<br />

Terminal.exe. The Terminal program will show an empty window.<br />

9. In the pull down menu, select "File" and than "Open". In the file<br />

selection box double click "Adj93.trm" to start the program: 2 lines of<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 53 of 68


function keys are now accessible at the bottom of the window.<br />

The upper line contains 5 keys, the lower line 4 keys plus a display of<br />

the real time clock. By loading the setup file "Adj93.trm" some of the 9<br />

function keys are assigned to a direct command of the autosampler.<br />

The function key "Level: x" (x can be 1,2,3 or4) allows to change the<br />

assignment of the function keys. When you click on this key, the<br />

program will change over to the next level and will offer the next set of<br />

8 direct commands. The Terminal program provides 4 levels with max.<br />

32 direct command assignments. Between the menu bar and the<br />

function keys is an area supposed to display the communication<br />

between the PC/Terminal.exe and the <strong>AS</strong>-<strong>93plus</strong>.<br />

10.Click on the function key "Get ID" of the set "Level 1". The terminal<br />

program will display the following window:<br />

The the first line of the communication area shows the direct command that has been sent<br />

to the autosampler. The second line shows the response of the autosampler sent back to<br />

the PC.<br />

A description of all direct commands are provided in chapter "Firmware Description"<br />

starting on page 64<br />

11.Click on function key "Level 1" in order to get another set of direct<br />

commands. With these commands the precision of the sample<br />

location access can be determined.<br />

12.Click on the function key "Load Tray F", to load the tray definition file<br />

for the sample tray group F.<br />

The commands on level 2 are assigned for sample tray type F, which<br />

is the standard set of sample racks. In this sample tray group the<br />

corner locations are 0, 5, 146 and 157.<br />

54 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


If you are using another sample tray type, the corner locations will be<br />

different, e.g. 0, 6, 186 and 200 for sample tray type E,.<br />

In this case you must edit the function keys of level 2 for the new<br />

locations as follows:<br />

1. In the pull down menu, click on "Settings", in "Settings" click on<br />

"Function Keys...".<br />

2. Make any corner location changes that you require<br />

3. Modify the load tray command and click "OK"<br />

(Load Tray E: set tray=4<br />

Load Tray F: set tray=5<br />

Load Tray G: set tray=6).<br />

The Terminal program will run now for an other sample tray group.<br />

13.Click on the function keys "Goto 0" and than "Z Down" to move the<br />

sampling probe holder over the rinsing port.<br />

14.Insert the sampling probe and check if it access exact the center of the<br />

rinsing port. If not, measure the distance between the sampling probe<br />

and the center of the rinsing port with a Millimeter scale in X- and Ydirection.<br />

Note the values in a table.<br />

15.Repeat this procedure accordingly for all corner positions.<br />

The table below shows an example of a misalignment:<br />

Table 4 Example of mechanical misalignment<br />

Sample location 0 5 146 157<br />

X-offset [mm] -1.0 -2.0 -1.0 -2.0<br />

Y-offset [mm] -2.0 -2.0 -2.0 -2.0<br />

Note A tolerance of ±1 mm is within specification<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 55 of 68


Kinds of positional misalignments<br />

There are 2 kinds of positional misalignment possible:<br />

• Parallel misalignment, if<br />

the positionings, seen in X-axis, of the rear sample locations<br />

(0,14,26,38,50,62,74,86,98,110,122,134,146) are ok,<br />

but the front sample locations<br />

(5,13,25,37,49,61,73,85,97,109,121,133,145,157) are showing an offset.<br />

• Offset misalignment,<br />

if the tip of the sampling probe shows a constant offset in X- or Y-axis in all<br />

sample locations .<br />

Note A parallel misalignment must be corrected before an offset misalignment<br />

can be corrected.<br />

The table 4 on page 55 shows the existence of a parallel misalignment while the X-offset<br />

is just in tolerance.<br />

Correction of a parallel misalignment<br />

A parallel misalignment must be corrected by rotating the tower.<br />

0<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

14<br />

15<br />

16<br />

17<br />

18<br />

19<br />

20<br />

21<br />

22<br />

23<br />

24<br />

25<br />

26<br />

27<br />

28<br />

29<br />

30<br />

31<br />

32<br />

33<br />

34<br />

35<br />

36<br />

37<br />

38<br />

39<br />

40<br />

41<br />

42<br />

43<br />

44<br />

45<br />

46<br />

47<br />

48<br />

49<br />

50<br />

51<br />

52<br />

53<br />

54<br />

55<br />

56<br />

57<br />

58<br />

59<br />

60<br />

61<br />

62<br />

63<br />

64<br />

65<br />

66<br />

67<br />

68<br />

69<br />

70<br />

71<br />

72<br />

73<br />

Figure 26 Correction of a parallel misalignment ( exaggerated illustration)<br />

56 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

74<br />

75<br />

76<br />

77<br />

78<br />

79<br />

80<br />

81<br />

82<br />

83<br />

84<br />

85<br />

86<br />

87<br />

88<br />

89<br />

90<br />

91<br />

92<br />

93<br />

94<br />

95<br />

96<br />

97<br />

98<br />

110 122 134 146<br />

99 111 123 135 147<br />

100 112 124 136 148<br />

101 113 125 137 149<br />

102 114 126 138 150<br />

103 115 127 139 151<br />

104 116 128 140 152<br />

105 117 129 141 153<br />

106 118 130 142 154<br />

107 119 131 143 155<br />

108 120 132 144 156<br />

109 121 133 145 157


1. Switch of the <strong>AS</strong>-93 by unplugging the power supply.<br />

2. Remove the sampling probe, the sample tubes and the sample racks<br />

from the sample tray table.<br />

3. With a knife carefully lift and remove the 4 caps that cover the<br />

mounting screws of the sample tray table.<br />

4. Remove the sample tray table and place it in front of the chassis.<br />

5. Loosen the 4 screws of the tower's bottom plate.<br />

6. Carefully turn the tower as much as the correction requires. Don't<br />

touch the arm during this procedure.<br />

7. To check the correction, mount the sample tray table and insert the left<br />

sample rack.<br />

8. Insert the sampling probe and move the sampling probe holder to<br />

location 0 and 5 by means of the Terminal program.<br />

9. Measure the X-offset with a millimeter scale. If there is still a different<br />

offset in X-direction repeat step 4 and 5.<br />

In general, a positional test can also preformed with the spectrometer<br />

operating software by means of the manual commands in the<br />

autosampler page.<br />

10.Check the X-offset also at the right side, i.e. position 146 and 157. If<br />

the X- offset on all corner locations is constant, a parallel misalignment<br />

is corrected.<br />

Figure 27 Location of the Tower Bottom Plate / Tower Mounting Screws<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 57 of 68


Correction of a X/Y-offset misalignment<br />

A X/Y-offset misalignment is present, if the sampling probe does not go exactly into the<br />

center of each position of the sample tray.<br />

Note A tolerance of ± 1 mm is allowed!<br />

The X/Y-offset misalignment will be corrected by entering offset steps for the X- and<br />

Y-drive. These offset steps are used for all movements as soon as they are stored in the<br />

non-volatile RAM area .<br />

The offset steps can be entered by using the utility Terminal.exe with the associated file<br />

Adj93.trm .<br />

1. Measure the precision of positioning as described in chapter<br />

Alignment using TERMINAL.EXE and ADJ93.TRM on page 52 by<br />

using the Terminal program, commands level 2, and note the values<br />

into a table. The table below shows an example:<br />

Table 5 Example of X/Y-Offset Misalignment<br />

Location X-Offset [mm] Y-Offset [mm]<br />

0 -1.0 -2.0<br />

5 -1.0 -2.0<br />

146 -1.0 -2.0<br />

157 -1.0 -2.0<br />

2. Click on the function key "Level 2" to make the following commands<br />

accessible:<br />

58 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


3. Click on the function key "Get X-Offset (mm)" and read the current xoffset<br />

stored in the non-volatile RAM.<br />

4. Calculate the new X-offset:<br />

The measured X-value is -1.0 mm, that means the sample probe is 1.0<br />

mm too far left and needs to be moved 1.0 mm to the right side.<br />

5. Click on the function key "Set X-Offset (mm)".<br />

Type in the calculated value "1.00" and press the "Enter" key.<br />

6. Click on the function key "Enable X-Offset".<br />

Give enough reaction time the autosampler between each new<br />

command input.<br />

7. Click on the function key "Initialize".<br />

8. Check the new precision in X-direction by using the level 2<br />

commands. If the new precision is not inside a tolerance of ± 1.0 mm<br />

repeat the steps 2-7.<br />

9. The correction of a Y-misalignment works similar as the above<br />

described X-alignment (step 2 until 8) by means of the "Y" function<br />

keys.<br />

Note The left end position of the tower is limited by the position of the hall<br />

sensor. X-offset steps (=mm) are allowed in the range of 0 to10 and will move the<br />

tower only from left to the right side. A negative number is not allowed. If a<br />

correction to the left side is required, the main board must be moved (see<br />

Figure 28).<br />

Note The rear end position of the sampling probe holder is limited by a optical<br />

sensor. Y-offset steps (=mm) are allowed in the range of 0 to10 and will move the<br />

sampling probe holder only from back to front. A negative number is not allowed.<br />

If a correction to the back will be required, the flag of the optical sensor must be<br />

moved. Take care that the sampling probe holder is in the sample location 0<br />

before you perform this alignment (see Figure29).<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 59 of 68


Figure 28 X-Offset Correction by Moving the PCB<br />

Figure 29 Adjustable Flag and Light Barrier for Y-Offset Alignment<br />

60 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Checking the Event Relay<br />

The Event relay is a potential-free relay contact that can be used to control an external<br />

system.<br />

Internal Cyclic tests<br />

1. Connect an Ohm meter to the terminal "Event". See item 2 of Electrical<br />

connectors and switches on page 15.<br />

2. Click on the function key "Event ON" of the level 4 commands.<br />

The relay contact will close the contact and the ohm-meter should<br />

show zero ohm. (You might also hear a "click" if the environmental<br />

noise level is not too high.)<br />

3. Click on the function key "Event OFF".<br />

The relay contact will be open and the ohm meter should show "high<br />

resistance".<br />

4 internal test programs have been build-in allowing to test the <strong>AS</strong>-<strong>93plus</strong> without a need<br />

to attach a PC. These "stand-alone tests" can be activated by setting the dipswitch 7 to<br />

ON using the dipswitch configurations as follows:<br />

1. Cyclic sample location test of sample tray group F.<br />

Continuous accessing of all sample location.<br />

2. Cyclic X-drive test<br />

The tower will move to the left and right end position.<br />

3. Cyclic Y-drive test.<br />

The sampling probe holder will move to the front and rear end position.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 61 of 68


4. Cyclic Z-drive test<br />

The arm will move up and down to the end position.<br />

5. Cyclic pump and movement test<br />

The arm will move to any position of the tray type "F" and peristaltic<br />

pump will run continuously.<br />

6. Cyclic pump test<br />

The peristaltic pump will run continuously.<br />

62 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Maintenance<br />

There are not much parts to maintain inside of the <strong>AS</strong>-<strong>93plus</strong>. Maintenance work is<br />

basically related to guide rods and spindles of the X-, Y- and Z-drive which should be kept<br />

slightly lubricated with a thin film of the following lubricants:<br />

1. Spindles:<br />

"Losymol" oil, P/N B0508429, comes in a 10 mL bottle<br />

2. Guide rods:<br />

"Nominol SD" grease, P/N B050-8430, comes in a 1 kg container<br />

Dust and/or the aggressive vapors of the lab environment can make the surface of the<br />

guide rods and spindles sticky over the time. In this case it might be necessary to clean<br />

those surfaces by means of a tissue and eventually an organic solvent prior of applying<br />

lubricants. Be careful with the use of Acetone. Acetone is an excellent solvent to remove<br />

grease and/or oil from metallic surfaced, but is also rather aggressive to plastic surfaces.<br />

It is recommended to protect somehow neighboring parts of the guide rods and spindles<br />

such as plastic parts or PCBs prior of cleaning surfaces and applying lubricants.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 63 of 68


Firmware<br />

The firmware of the <strong>AS</strong>-<strong>93plus</strong> is identical with the firmware of the <strong>AS</strong>-90plus. During<br />

power-on initialization the presence of the pump module will be checked as a procedure<br />

of self-identification. The autosampler will be initialized as an <strong>AS</strong>-<strong>93plus</strong>, if the pump is<br />

connected to the controller PCB. Otherwise the firmware will initialize as an <strong>AS</strong>-90plus.<br />

The firmware is stored in a FL<strong>AS</strong>H-RAM located on the controller PCB and can be<br />

downloaded via the <strong>AS</strong>-<strong>93plus</strong> Utilities program. A procedure is given in chapter Firmware<br />

Download on page 50.<br />

Direct commands<br />

The control of the autosampler is done usually by means of direct commands sent from<br />

the application software / Device Control Module (DCM) to the mechanical unit. It is the<br />

purpose of the firmware to interpret the command and convert them into action. In addition<br />

the firmware has to monitor all sensor signal and has to send data back to the application.<br />

For trouble-shooting purpose it might be useful to have a list of direct commands available<br />

as shown below:<br />

Identification Commands<br />

• get id query<br />

Autosampler identification name<br />

• get syst.vers query<br />

firmware version with date and time<br />

• get syst.serialno query<br />

Serial number of the autosampler<br />

• get syst.part query<br />

Part number of the autosampler<br />

Reset Command<br />

• reset action<br />

only<br />

Power on reset. To achieve the same effects as switching power on.<br />

Tray commands<br />

• set tray= action<br />

Load type of sample tray<br />

Parameter: 4,5,6 (tray E=4, tray F=5, tray G=6)<br />

• get tray query<br />

Read loaded type of sample tray<br />

64 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


Pump Commands<br />

• stop control.pump action<br />

Switch off the pump<br />

• start control.pump action<br />

Switch on the pump<br />

X-, Y-, Z- Drive commands<br />

• init cntrl.x action<br />

X-drive moves to the home position<br />

• init cntrl.y action<br />

Y-drive moves to the home position<br />

• init cntrl.z action<br />

Z-drive moves to the home position<br />

• set up= action<br />

Move the autosampler arm upwards to position <br />

Parameter: 0.0-146.0,<br />

Default: 0<br />

• get up query<br />

Read the current z-drive position<br />

• set down= action<br />

Move the autosampler arm downwards to position for a<br />

duration of milliseconds<br />

Parameter a: 0.0-146.0,<br />

Parameter b: no range check (resolution 15,625 ms), 0= stay forever<br />

Default: 0,0<br />

• get down query<br />

Read the current z-drive position<br />

• set posup= action<br />

Move the z-drive to the upper end-position, then move x- and y-drive to<br />

sample position . Z-drive doesn’t move downwards<br />

Parameter: Tray E=0-200, Tray F=0-157, Tray G=0-62<br />

• get posup query<br />

Read current sample position<br />

• get cntrl.x.soff query<br />

Read the current X-offset in mm<br />

• set cntrl.x.soff _ action<br />

Enter new X-offset in mm<br />

parameter: 0-10<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 65 of 68


• get cntrl.y.soff query<br />

Read the current Y-offset in mm<br />

• set cntrl.y.soff action<br />

Enter new Y-offset in mm<br />

parameter: 0-10<br />

• enable cntrl.x action<br />

Save new X-offset value into FL<strong>AS</strong>H<br />

• enable cntrl.y action<br />

Save new Y-offset value into FL<strong>AS</strong>H<br />

Event Commands<br />

• set cntrl.event.test action<br />

Switch on/off the event relay<br />

Parameter: on,off<br />

66 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>


<strong>AS</strong>-<strong>93plus</strong> Error Code Reference Table<br />

Code Description<br />

0 No error<br />

2 Parameter error<br />

Table 6 <strong>AS</strong>-<strong>93plus</strong> Error Code Table<br />

3 Value out of range<br />

7 Command error<br />

9 EPCI error<br />

21 No autosampler connected<br />

100 Autosampler time out<br />

101 Autosampler X motor time out<br />

102 Autosampler y motor time out<br />

103 Autosampler Z motor time out<br />

105 Illegal X-Y coordinates<br />

106 Up move out of range<br />

107 Down move out of range<br />

109 Autosampler position out of range<br />

110 Illegal liquid detector status<br />

111 No liquid detected<br />

121 Checksum error<br />

122 Reset task execution error<br />

130 S-Record error<br />

131 S-Record checksum error<br />

132 Transmission error<br />

133 Synchronization error<br />

134 No start of transmission error<br />

135 No end of transmission error<br />

136 Header checksum error<br />

137 Data checksum error<br />

200 Autosampler time out<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 67 of 68


Code Description<br />

Table 6 <strong>AS</strong>-<strong>93plus</strong> Error Code Table<br />

201 Autosampler busy error<br />

202 Motor time out error<br />

203 Motor initialization error<br />

204 Motor initialization error<br />

205 Motor break error<br />

301 X motor time out error<br />

302 X motor initialization error<br />

304 X motor break error<br />

311 Y motor time out error<br />

312 Y motor initialization error<br />

314 Y motor break error<br />

321 Z motor time out error<br />

322 Z motor initialization error<br />

326 Z motor liquid detector error<br />

330 Illegal X-Y coordinates<br />

331 Up move out of range<br />

332 Down move out of range<br />

333 Autosampler position out of range<br />

68 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!