Fitting pump tubes to the rinse pump<br />

1. Swing the cam lever fully<br />

clockwise to release the<br />

pressure block.<br />

2. Pull the pressure block off<br />

the pivot.<br />

3. Fit the rinse feed pump<br />

tube:<br />

1. Hook a collar into the<br />

inside channel of a<br />

retaining hook, using<br />

the correct slot for the<br />

diameter of the tube.<br />

2. Gently tension the pump<br />

tube around the pump<br />

rotor and hook the<br />

second collar into the<br />

inside channel of the<br />

other retaining hook.<br />

Do not twist or kink the<br />

pump tube.<br />

4. Fit the drain pump tube:<br />

1. Repeat substeps 1 and<br />

2, above, to fit the drain<br />

pump tube to the<br />

outside channel of the<br />

retaining hooks.<br />

Collars<br />

5. Place a drop of silicone oil on the pump tubes in contact with the pump<br />

rollers to reduce wear.<br />

6. Slide the pressure block back onto the pivot.<br />

Retaining hook<br />

Retaining hook<br />

Cam lever<br />

Pressure<br />

block<br />

7. Swing the cam lever anti-clockwise to press the pressure block<br />

against the pump tubes.<br />

The final position of the cam lever depends on the pressure that is<br />

required on the pump tubes for satisfactory operation (the “3 o’clock”<br />

position is usually about right). Check the flow of liquid through the<br />

rinsing port when the pump is operating. If the flow is intermittent or<br />

inadequate, increase the pressure on the pump tubes with the cam<br />

lever. The lifetime of the pump tubes will be shortened if the pressure<br />

is too high.<br />

Note: Release the pressure on the pump tubes during longer pauses in operation (e.g.<br />

overnight) to prolong the lifetime of the pump tubes.<br />

Make sure that the inner diameter of the drain pump tube is the same as, or larger than,<br />

the rinse feed pump tube. If the drain pump tube has a smaller inner diameter, rinse<br />

liquid will back up in the rinsing port and ultimately overflow into the autosampler,<br />

possibly causing damage.<br />

38 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />

Pivot<br />

Pump tube<br />

Pump<br />

rotor