Fitting pump tubes to the rinse pump<br />
1. Swing the cam lever fully<br />
clockwise to release the<br />
pressure block.<br />
2. Pull the pressure block off<br />
the pivot.<br />
3. Fit the rinse feed pump<br />
tube:<br />
1. Hook a collar into the<br />
inside channel of a<br />
retaining hook, using<br />
the correct slot for the<br />
diameter of the tube.<br />
2. Gently tension the pump<br />
tube around the pump<br />
rotor and hook the<br />
second collar into the<br />
inside channel of the<br />
other retaining hook.<br />
Do not twist or kink the<br />
pump tube.<br />
4. Fit the drain pump tube:<br />
1. Repeat substeps 1 and<br />
2, above, to fit the drain<br />
pump tube to the<br />
outside channel of the<br />
retaining hooks.<br />
Collars<br />
5. Place a drop of silicone oil on the pump tubes in contact with the pump<br />
rollers to reduce wear.<br />
6. Slide the pressure block back onto the pivot.<br />
Retaining hook<br />
Retaining hook<br />
Cam lever<br />
Pressure<br />
block<br />
7. Swing the cam lever anti-clockwise to press the pressure block<br />
against the pump tubes.<br />
The final position of the cam lever depends on the pressure that is<br />
required on the pump tubes for satisfactory operation (the “3 o’clock”<br />
position is usually about right). Check the flow of liquid through the<br />
rinsing port when the pump is operating. If the flow is intermittent or<br />
inadequate, increase the pressure on the pump tubes with the cam<br />
lever. The lifetime of the pump tubes will be shortened if the pressure<br />
is too high.<br />
Note: Release the pressure on the pump tubes during longer pauses in operation (e.g.<br />
overnight) to prolong the lifetime of the pump tubes.<br />
Make sure that the inner diameter of the drain pump tube is the same as, or larger than,<br />
the rinse feed pump tube. If the drain pump tube has a smaller inner diameter, rinse<br />
liquid will back up in the rinsing port and ultimately overflow into the autosampler,<br />
possibly causing damage.<br />
38 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong><br />
Pivot<br />
Pump tube<br />
Pump<br />
rotor