Connecting the sample tube<br />

The sample uptake rate is governed by the internal diameter of the sample tube. Two<br />

sample tubes with different internal diameters are provided with the autosampler. Each<br />

sample tube has a screw connector on one end for connection to the sampling probe and<br />

a screw connector on the other end for connection to the nebulizer via an adapter.<br />

You use the sample tube with 0.6mm internal diameter (B300-0158) for routine samples.<br />

You use the sample tube with 1mm internal diameter (B019-1060) if you are operating<br />

the burner system with a High Performance Nebulizer; this sample tube allows the<br />

nebulizer to attain its maximum performance.<br />

To connect the sample tube to the sampling probe<br />

Tower cover<br />

.<br />

Depending on how you position the autosampler with respect to the rest of the system,<br />

you can attach the sample tube to either the front of the autosampler arm or the rear of<br />

the tower cover. A spring clip with two slots is provided to hold the sample tube; the<br />

narrower slot holds the 0.6 mm tube, while the wider slot holds the 1mm tube.<br />

1. Screw the connector on the sample tube finger-tight into the fitting on<br />

the sampling probe.<br />

2. Make a smooth loop in the sample tube and then attach it to either the<br />

front of the autosampler arm or the rear of the tower cover:<br />

1. Insert the spring clip into the appropriate socket.<br />

2. Thread the sample tube gently around the hook in the correct slot.<br />

Note Do not kink or pinch the sample tube.<br />

Spring clip<br />

with two slots<br />

Note The loop must be large enough so that the sampling probe holder can<br />

move unhindered backward and forward along the autosampler arm.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 21 of 68