AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...
AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...
AS9N010B.STB AS-93plus Service Manual.fm - LIQUID-scan GmbH ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Checking the Event Relay<br />
The Event relay is a potential-free relay contact that can be used to control an external<br />
system.<br />
Internal Cyclic tests<br />
1. Connect an Ohm meter to the terminal "Event". See item 2 of Electrical<br />
connectors and switches on page 15.<br />
2. Click on the function key "Event ON" of the level 4 commands.<br />
The relay contact will close the contact and the ohm-meter should<br />
show zero ohm. (You might also hear a "click" if the environmental<br />
noise level is not too high.)<br />
3. Click on the function key "Event OFF".<br />
The relay contact will be open and the ohm meter should show "high<br />
resistance".<br />
4 internal test programs have been build-in allowing to test the <strong>AS</strong>-<strong>93plus</strong> without a need<br />
to attach a PC. These "stand-alone tests" can be activated by setting the dipswitch 7 to<br />
ON using the dipswitch configurations as follows:<br />
1. Cyclic sample location test of sample tray group F.<br />
Continuous accessing of all sample location.<br />
2. Cyclic X-drive test<br />
The tower will move to the left and right end position.<br />
3. Cyclic Y-drive test.<br />
The sampling probe holder will move to the front and rear end position.<br />
<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 61 of 68