Checking the Event Relay<br />

The Event relay is a potential-free relay contact that can be used to control an external<br />

system.<br />

Internal Cyclic tests<br />

1. Connect an Ohm meter to the terminal "Event". See item 2 of Electrical<br />

connectors and switches on page 15.<br />

2. Click on the function key "Event ON" of the level 4 commands.<br />

The relay contact will close the contact and the ohm-meter should<br />

show zero ohm. (You might also hear a "click" if the environmental<br />

noise level is not too high.)<br />

3. Click on the function key "Event OFF".<br />

The relay contact will be open and the ohm meter should show "high<br />

resistance".<br />

4 internal test programs have been build-in allowing to test the <strong>AS</strong>-<strong>93plus</strong> without a need<br />

to attach a PC. These "stand-alone tests" can be activated by setting the dipswitch 7 to<br />

ON using the dipswitch configurations as follows:<br />

1. Cyclic sample location test of sample tray group F.<br />

Continuous accessing of all sample location.<br />

2. Cyclic X-drive test<br />

The tower will move to the left and right end position.<br />