7. Click on "Next" to finish the reconfiguration procedure.<br />
8. Double click the Icon "AAWinLab Analyst" to start the reconfigured<br />
spectrometer operating software. After a while the AAWinLab’s initial<br />
screen will be shown with it’s menu line and toolbar.<br />
9. Select "File" - "New" to create a method to test the function and<br />
precision of the X- and Y-drive of the <strong>AS</strong>-<strong>93plus</strong>.<br />
10.Select an element (e.g. Cu) from the "Create New Method" window<br />
and click "OK".<br />
11.Type in the method description. Then select the tab "Calib." and click<br />
on the sub-tab "Autosampler".<br />
Figure 16 Autosampler Window<br />
12.Select the "Tray Type" (rack group). and select "F" as this is the<br />
standard rack configuration of the <strong>AS</strong>-<strong>93plus</strong>.<br />
Note Figure 14 shows the 3 types of predefined tray types (=rack groups) using<br />
Perkin Elmer racks. The character E,F and G of the tray type designation<br />
identifies the rack group while the following numbers points out the number of the<br />
maximal sample locations of the tray type. Refer to page 25 for details about predefined<br />
"tray types" = rack groups.<br />
If you are using sample racks from other suppliers you have to copy appropriate<br />
"tray files" that are named according the 8.3 character file name convention into<br />
the AA Winlab/Elan directory. "Ready-to-use" tray -files are provided with the <strong>AS</strong>-<br />
93 utility disk or must be prepared as described on page 47.<br />
<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 41 of 68