Tubing configurations<br />

You can set up several different tubing configurations to meet the given requirements.<br />

• Configuration 1. The rinse pump is used to pump rinse liquid into the rinsing port<br />

and to pump the liquid back out into the waste bottle. A narrow-bore pump tube<br />

(B314-0730; red-red) is used for the rinse feed to provide a slow flow rate of the<br />

rinse liquid. This is the preferred configuration for ICP-OES and ICP-MS<br />

applications.<br />

• Configuration 2. The rinse pump is used to pump rinse liquid into the rinsing port<br />

and to pump the liquid back out into the waste bottle. A wide-bore pump tube<br />

(B314-0721; violet-white) is used for the rinse feed to provide a fast flow rate of<br />

the rinse liquid. This is the preferred configuration for flame A<strong>AS</strong> applications.<br />

• Configuration 3. The rinse pump is used to pump rinse liquid into the rinsing port<br />

only. The drain tube is connected to the rinsing port directly and the liquid drains<br />

by gravity into the waste bottle. A wide-bore pump tube (B314-0721; violetwhite)<br />

is used for the rinse feed to provide a fast flow rate of the rinse liquid.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 31 of 68