Correction of a X/Y-offset misalignment<br />

A X/Y-offset misalignment is present, if the sampling probe does not go exactly into the<br />

center of each position of the sample tray.<br />

Note A tolerance of ± 1 mm is allowed!<br />

The X/Y-offset misalignment will be corrected by entering offset steps for the X- and<br />

Y-drive. These offset steps are used for all movements as soon as they are stored in the<br />

non-volatile RAM area .<br />

The offset steps can be entered by using the utility Terminal.exe with the associated file<br />

Adj93.trm .<br />

1. Measure the precision of positioning as described in chapter<br />

Alignment using TERMINAL.EXE and ADJ93.TRM on page 52 by<br />

using the Terminal program, commands level 2, and note the values<br />

into a table. The table below shows an example:<br />

Table 5 Example of X/Y-Offset Misalignment<br />

Location X-Offset [mm] Y-Offset [mm]<br />

0 -1.0 -2.0<br />

5 -1.0 -2.0<br />

146 -1.0 -2.0<br />

157 -1.0 -2.0<br />

2. Click on the function key "Level 2" to make the following commands<br />

accessible:<br />

58 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>