Correction of a X/Y-offset misalignment<br />
A X/Y-offset misalignment is present, if the sampling probe does not go exactly into the<br />
center of each position of the sample tray.<br />
Note A tolerance of ± 1 mm is allowed!<br />
The X/Y-offset misalignment will be corrected by entering offset steps for the X- and<br />
Y-drive. These offset steps are used for all movements as soon as they are stored in the<br />
non-volatile RAM area .<br />
The offset steps can be entered by using the utility Terminal.exe with the associated file<br />
Adj93.trm .<br />
1. Measure the precision of positioning as described in chapter<br />
Alignment using TERMINAL.EXE and ADJ93.TRM on page 52 by<br />
using the Terminal program, commands level 2, and note the values<br />
into a table. The table below shows an example:<br />
Table 5 Example of X/Y-Offset Misalignment<br />
Location X-Offset [mm] Y-Offset [mm]<br />
0 -1.0 -2.0<br />
5 -1.0 -2.0<br />
146 -1.0 -2.0<br />
157 -1.0 -2.0<br />
2. Click on the function key "Level 2" to make the following commands<br />
accessible:<br />
58 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>