Setting up the rinsing port<br />

Connecting the tubing<br />

Pump tubes and retaining hooks<br />

The pump tubes have three collars, giving two sections that can be tensioned around the<br />

pump rotor. If one section of a tube becomes deformed, and there is no other damage,<br />

you can use the other section.<br />

Pump tubes of different internal diameter are available so that you can vary the flow rate<br />

of the rinse liquid through the rinsing port. The pump tubes have colored collars for easy<br />

identification.<br />

Pump tube<br />

Color<br />

(collars)<br />

The retaining hooks on the rinse pump have an inside channel for the rinse feed pump<br />

tube and an outside channel for the drain pump tube (see Figure 9). There are slots for the<br />

pump tubes on the top and bottom of each retaining hook; the slots on the top are for tubes<br />

with narrower diameter, while the slots on the bottom are for tubes with wider diameter.<br />

Figure 9 Rinse Pump<br />

i.d. (mm)<br />

B314-0730 red-red 1.14 3.5–5<br />

B314-0721 violet-white 2.79 11–14<br />

Rear channel<br />

Front channel<br />

Pump rotor<br />

Section 1 Section 2<br />

Top view<br />

Collars Pump tube<br />

Retaining hook<br />

Nominal flow<br />

rate (mL/min)<br />

Side view<br />

Retaining hook<br />

Retaining hook<br />

Pump<br />

rotor<br />

30 of 68 <strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong>