Connecting the tubes for configuration 2<br />

1. Fit the rinse feed pump tube (violet-white) (5) and the drain pump tube<br />

(violet-white) (3) to the rinse pump as described in Fitting pump tubes<br />

to the rinse pump on page 38.<br />

Note Always fit a wide-bore drain pump tube (violet-white).<br />

2. Connect up the rinse feed line:<br />

1. Push a barbed fitting of the tube connector (4) into<br />

the PVC tube prefitted to the inlet fitting of the<br />

rinsing port.<br />

2. Push the end of the rinse feed pump tube ( 5) coming<br />

from the upper retaining hook onto the other barbed<br />

fitting of the tube connector (4).<br />

3. Push the barbed fitting of adapter C (7) into the rinse<br />

feed pump tube coming from the lower retaining<br />

hook.<br />

4. Screw the connector on the rinse liquid feed tube (8)<br />

into adapter C.<br />

Tube connector<br />

Adapter<br />

3. Connect up the drain line:<br />

1. Screw adapter M (2) into adapter E (1).<br />

2. Push the barbed fitting of adapter E into the silicone tube prefitted<br />

to the drain nozzle on the rinsing port.<br />

3. Push the end of the drain tube (3) coming from the lower retaining<br />

hook onto the barbed fitting of adapter M.<br />

4. Push the tube connector (6) into the end of the drain pump tube<br />

coming from the upper retaining hook.<br />

5. Push one end of the PVC drain tube (9) onto the connector.<br />

<strong><strong>AS</strong>9N010B</strong>.<strong>STB</strong> 35 of 68