6 OperationIf you want to start the definition of the cold resistance, proceed as described in chapter 5.1and confirm a change of the sensor or start from the main menu with (see chapter 5.2).1)WARNING!Watch the sensor’s temperature!Proceed the definition of the cold resistance only, if your sensor has cooled down to ambienttemperature.If the sensor is too warm, you will get a wrong result and the sensor may be damaged.Let the sensor cool down <strong>for</strong> at least 5 min. A longer cool down period will be better.Check the your sensor’s temperature.2)NOTICE!Protect your sensor from air drought.If your sensor is exposed to air draught during the definition of the cold resistance, this mayfalsify the result or the measurement fails.Take care <strong>for</strong> a air draught free environment.Press keyto end the definition of the cold resistance. The installation starts.32