12.07.2015 Views

DESIGN AND DEVELOPMENT OF MEDICAL ELECTRONIC ...

DESIGN AND DEVELOPMENT OF MEDICAL ELECTRONIC ...

DESIGN AND DEVELOPMENT OF MEDICAL ELECTRONIC ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ELECTRODE 1ELECTRODE 2I_OUTR3+15V_SWITCHED1M +15V_SWITCHEDR21MR125kR41MC30.01uFC40.01uF74268C60.01uF+5V_SWITCHEDIC1ICM7555/TO5 3CONTVO UTDISCHRESETTRIGTHOLDV+1+5V416132JKCLKCLRPREIC2A74C76IC3AQQ1 274C04151412R5100kR7100kR11100kIC5UPC4250C70.01uF3245IC4A74C08IC4B74C08367+-4 86R91MC20.01uF-15V_ SWITCHEDR62MR102MC11uF+C51uF+327+-4 8R81MGND-15V_SWITCHEDR12IC6UPC42506+5V_SWITCHEDR131M+15V+15V_SWITCHED100k-15V-15V_SWITCHED+5V+5V_SWITCHED30s to 4min30s to 4min30min to 20hoursFigure 7.12 In this symmetric biphasic current stimulator, pulses generated by timer IC1 are reshaped by IC2A, IC3A, IC4A, IC4B, andIC5 into a symmetric bipolar square-pulse train. The resulting voltage waveform is converted into a constant-current waveform by the invertingHowland current pump built around IC6. Livnat et al. [1981] used implanted electrodes and this output stage for chronic stimulation ofthe central nervous system.323