Q17IRF9530+24V_SW+15V_NON_ISOGND NON ISO1IC182 3+VIN +VOUT-VINCOMM-VOUTHBO4U15D15 DC-DC Converter4512BPSGIC43BNX002CBCG43D1333V, 5WQ39TIP42R331.0IC19UC3906D12SCHOTTKY16Q19TIP 421512R106100K11R35100K+24_VBATTO BATTERYR104470KBATT_COMMQ18BS170R10510M+24V_ENABLEC22470uF, 50V1+IN2IC20DSA6U48S54+OUT3-IN-OUTSINK243579C/S -C/LC/S ++VINPWR INDOC INDC260.22uF14 1 8 6C270.1uFVOLT. SENSESTATE LEVEL CNTL1310R39100KR36385KR379.09KC2310uFC24.01uFC250.1uFR381MQ20IRF9530+5V+5SWCOMPC/S OUTOC TERMGNDSOURCECHRG ENABLET. BIAS+5V_ENABLEGNDFigure 8.37 An isolated dc/dc rated at 3 kV with a maximum 60-Hz leakage of a 2-µA converter powers the battery charger circuitry. Asingle-chip gel-cell charge controller IC (UC3906) charges two sealed 12-V batteries. Q4 switches power to the high-voltage power supplyfrom the 24-V battery. This switch is turned-on by Q5 upon receipt of the appropriate command from the shock-box microcontroller. IC9produces 5 V to run the microcontroller and associated logic.422