I1TR280IR_TISSUE100k+ –d/dt d/dt10uPOTR24IC =1350V+C22100uFR740mC32100uFR840mR6330kR10330kVCAPACITOR BANKC92100uFR1240mC52100uFR1340mR18330kR10330kC102100uFR1540mC72100uFR1640mR19330kD14Dbreak-XD13Dbreak-XR17330kD9Dbreak-X+-V20 0R_COIL100 mHX2C228AITX11TR40C42100uFR940mR11330kC62100uFR2040mR21330kR2240mC82100uFR2240mR23330k0Figure 7.25 Suitable coil inductances for the magnetic stimulator can be evaluated using this simplified PSpice model. Since InternationalRectifier does not have a PSpice model available for the thyristor, we modified the instance model for a C228A SCR (which is totallydifferent from a ST330S16P0) to model the parameters of the ST330S16P0. Similarly, a breakaway diode model (DBREAK) was modifiedto simulate the behavior of the 40HF100 diodes. The simulation is used to to find an inductance that keeps the peak current within the SCR’scurrent-handling capability (8.3 kA) while conforming to the SCR’s switch-on capability (dI/dt 400 V/µs).V30+–5000350