ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls
ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls
ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
2.2 THE <strong>OXYGEN</strong> <strong>PROBE</strong><br />
The probe assembly provides a means of exposing the sensor to the atmosphere to be measured with sensor,<br />
thermocouple and heater wires connected via a weatherproof plug to the analyser lead. Reference air is fed via the<br />
plug for unheated probes and via a separate gas thread connection for heated probes.<br />
Connections are provided for an in-situ gas calibration check. A cleaning purge of air can be admitted via the cal.<br />
gas entry. The outer sheath can be metal or ceramic, depending on the application.<br />
In-situ zirconia oxygen probes will give a lower oxygen reading than a sampled gas measurement on a<br />
chromatograph or paramagnetic analyser because the flue gas contains a significant level of water vapour and a<br />
sampling system removes the water vapour through condensation. The oxygen content then appears as a higher<br />
percentage of the remaining gas. For example:<br />
If the gas contained five parts oxygen and fifteen parts moisture, removing the moisture would leave the oxygen at<br />
5.88%. This phenonema will depend on the fuel and the completeness of combustion. They are common to all<br />
zirconia oxygen sensors.<br />
Probes of 1000 mm normally have sufficient length for any installation. Customers requiring probes longer than<br />
1000 mm are supplied with a flow guide tube which uses the flue velocity to pull flue gas through a filter at the<br />
sensing tip and exhaust it near the flue wall.<br />
2.3 THE ANALYSER<br />
2.4 ALARMS<br />
The 1532 analyser is a transmitter with two 4–20 mA. outputs. One output is % oxygen with selectable span (Refer<br />
to Set-up Steps 4 and 5 in Section 5.5).<br />
The second output can be selected as oxygen deficiency, efficiency, flue temperature, reducing oxygen, percent<br />
carbon dioxide, probe EMF or a logarithmic oxygen range. Four alarm relays are provided. Refer to the<br />
specifications section for more details.<br />
The 1532 analyser is designed to operate with either a heated or unheated, zirconia probe. The analyser maintains<br />
the temperature of heated probes at 720°C. If the flue gas temperature is above 730°C, the probe heater cuts out<br />
completely and the process provides probe heating. The analyser solves the Nernst equation and will provide<br />
accurate oxygen measurements up to 1500°C (2730°F), although most probes are suitable only to 1400°C (2250°F)<br />
and some are limited to 900°C (1650°F).<br />
A block diagram of the analyser is shown in Figure 2.2 and details on its functions are given in the remainder of<br />
this section.<br />
2.5 HEATER SUPPLY<br />
Refer to OPERATOR FUNCTIONS Section for details on alarm functions.<br />
CAUTION<br />
The probe heater is supplied with 110 VAC at 1 A (cold). This supply has electrical shock danger to maintenance<br />
personnel. Always isolate the analyser before working with the probe. For maximum safety the heater supply is<br />
transformer isolated.<br />
2.2 NOVATECH <strong>OXYGEN</strong> ANALYSER