14.12.2012 Views

ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls

ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls

ZIRCONIA OXYGEN PROBE OUTPUT (mV) - Novatech Controls

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2.2 THE <strong>OXYGEN</strong> <strong>PROBE</strong><br />

The probe assembly provides a means of exposing the sensor to the atmosphere to be measured with sensor,<br />

thermocouple and heater wires connected via a weatherproof plug to the analyser lead. Reference air is fed via the<br />

plug for unheated probes and via a separate gas thread connection for heated probes.<br />

Connections are provided for an in-situ gas calibration check. A cleaning purge of air can be admitted via the cal.<br />

gas entry. The outer sheath can be metal or ceramic, depending on the application.<br />

In-situ zirconia oxygen probes will give a lower oxygen reading than a sampled gas measurement on a<br />

chromatograph or paramagnetic analyser because the flue gas contains a significant level of water vapour and a<br />

sampling system removes the water vapour through condensation. The oxygen content then appears as a higher<br />

percentage of the remaining gas. For example:<br />

If the gas contained five parts oxygen and fifteen parts moisture, removing the moisture would leave the oxygen at<br />

5.88%. This phenonema will depend on the fuel and the completeness of combustion. They are common to all<br />

zirconia oxygen sensors.<br />

Probes of 1000 mm normally have sufficient length for any installation. Customers requiring probes longer than<br />

1000 mm are supplied with a flow guide tube which uses the flue velocity to pull flue gas through a filter at the<br />

sensing tip and exhaust it near the flue wall.<br />

2.3 THE ANALYSER<br />

2.4 ALARMS<br />

The 1532 analyser is a transmitter with two 4–20 mA. outputs. One output is % oxygen with selectable span (Refer<br />

to Set-up Steps 4 and 5 in Section 5.5).<br />

The second output can be selected as oxygen deficiency, efficiency, flue temperature, reducing oxygen, percent<br />

carbon dioxide, probe EMF or a logarithmic oxygen range. Four alarm relays are provided. Refer to the<br />

specifications section for more details.<br />

The 1532 analyser is designed to operate with either a heated or unheated, zirconia probe. The analyser maintains<br />

the temperature of heated probes at 720°C. If the flue gas temperature is above 730°C, the probe heater cuts out<br />

completely and the process provides probe heating. The analyser solves the Nernst equation and will provide<br />

accurate oxygen measurements up to 1500°C (2730°F), although most probes are suitable only to 1400°C (2250°F)<br />

and some are limited to 900°C (1650°F).<br />

A block diagram of the analyser is shown in Figure 2.2 and details on its functions are given in the remainder of<br />

this section.<br />

2.5 HEATER SUPPLY<br />

Refer to OPERATOR FUNCTIONS Section for details on alarm functions.<br />

CAUTION<br />

The probe heater is supplied with 110 VAC at 1 A (cold). This supply has electrical shock danger to maintenance<br />

personnel. Always isolate the analyser before working with the probe. For maximum safety the heater supply is<br />

transformer isolated.<br />

2.2 NOVATECH <strong>OXYGEN</strong> ANALYSER

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!