5.5 SET-UP FUNCTION DETAILS<br />

1. SENSOR TYPE<br />

Options<br />

Model No. Enter the probe in use<br />

1. 1231 Heated Probe<br />

2. 1232 Unheated Probe<br />

3. 1233 Unheated Probe for high sulphur applications<br />

(cermet sheath)<br />

2. <strong>PROBE</strong> THERMOCOUPLE TYPE<br />

The probe can have either a type K, R, or N thermocouple as a sensor temperature detector.<br />

Options<br />

1. K Check in the manual Section 1<br />

2. R for the probe model number.<br />

3. N Enter the correct TC type.<br />

3. FLUE THERMOCOUPLE TYPE<br />

Select the flue thermocouple type.<br />

Options<br />

1. T<br />

2. J<br />

3. K<br />

4. R<br />

5. S<br />

6. N<br />

7. Nil<br />

NOTE<br />

For heated probes the flue thermocouple is a separate sensor from the oxygen probe and should be mounted near to<br />

and upstream from the probe. It is optional. If no thermocouple is required, select option ‘NIL’. In this case<br />

efficiency and flue temperature readouts will not be operable.<br />

With an unheated probe, the thermocouple within the probe may be used to sense flue temperature. In this case<br />

connect Terminals 3 and 4 to Terminals 5 and 6 with copper wire.<br />

4. TRANSMITTER <strong>OUTPUT</strong> CHANNEL 1<br />

Select the type of output required from Channel 1. Linear is the most common output required. The logarithmic<br />

output is often used where analog indicators give an exploded view of the oxygen range near stoichiometry. You<br />

can draw your own scale using data in Appendix 3.<br />

Options:<br />

1. Linear<br />

2. Logarithmic<br />

3. Reducing<br />

The reducing output is for special applications requiring extreme reducing conditions e.g. ceramic surface<br />

treatment.<br />

Linear output spans are adjustable in Set-up step 5. The logarithmic output is fixed at 0.1 to 20 % oxygen and the<br />

reducing output is fired at 10 -1 to 10 -25 % oxygen. If either of the latter two are selected, then set-up 5 will be<br />

skipped.<br />

NOVATECH <strong>OXYGEN</strong> ANALYSER 5.3