4.2 ALARM BUTTON<br />
Figure 4.1 Operator's Panel<br />
Repeatedly pressing the operators ‘ALARM’ button will produce alarm displays in sequence on the lower line of<br />
the LCD display. If an alarm has cleared prior to pressing the ‘ALARM’ button, it will not re-appear on a second<br />
run through the alarms. Active alarms which have been previously displayed will have ‘ACC’ (accepted),<br />
displayed alongside. New alarms will not have ‘ACC’ displayed until a second press of the ‘ALARM’ button.<br />
After the last active alarm is indicated, the lower line of the display will return to the last displayed lower line<br />
variable.<br />
The alarm ‘LED’ will flash on alarm. Pressing the ‘ALARM’ button will cause the LED to go steady if any alarms<br />
are still active, or extinguish if there are no active alarms. The horn relay will operate when an alarm occurs.<br />
Pressing ‘ALARM’ will mute the horn relay which will re-initiate on any new alarms.<br />
4.3 ALARM SCHEDULE<br />
4.3.1 SUMMARY OF ALARMS<br />
ALARM DESCRIPTION<br />
‘SENSOR FAIL’ Oxygen cell or electrode failure (high impedance); (inhibited under 650°C).<br />
‘<strong>PROBE</strong> HEATER In the first 20 minutes of power being applied to<br />
FAILURE’ the heater after being switched on, this alarm will not occur, but a ‘<strong>PROBE</strong><br />
UNDER–TEMPERATURE’ display will occur and relay RL8 will be activated.<br />
Refer to Section 6.13.<br />
‘<strong>PROBE</strong> TC O/C’ Probe thermocouple is open circuit. The heater in heated probes will switch off.<br />
‘REF AIR FAIL’ Low reference air flow to probe.<br />
‘FLOW SWITCH’ The reference air flow switch has failed. This is tested by automatically<br />
switching off the reference air pump once every ten minutes for three seconds to<br />
check the operation of the flow switch.<br />
‘A/D CAL ERROR’ The analog to digital converter has been found to fall outside the normal<br />
calibration specifications. This is an electronic fault.<br />
‘D/A CAL ERROR’ The digital to analog and voltage isolator circuit has been found to fall outside<br />
the normal calibration specifications. This check is only performed when the<br />
‘AUTO CAL’ button is pressed. Refer to Section 6.6.<br />
3.2 NOVATECH <strong>OXYGEN</strong> ANALYSER