4.2 ALARM BUTTON<br />

Figure 4.1 Operator's Panel<br />

Repeatedly pressing the operators ‘ALARM’ button will produce alarm displays in sequence on the lower line of<br />

the LCD display. If an alarm has cleared prior to pressing the ‘ALARM’ button, it will not re-appear on a second<br />

run through the alarms. Active alarms which have been previously displayed will have ‘ACC’ (accepted),<br />

displayed alongside. New alarms will not have ‘ACC’ displayed until a second press of the ‘ALARM’ button.<br />

After the last active alarm is indicated, the lower line of the display will return to the last displayed lower line<br />

variable.<br />

The alarm ‘LED’ will flash on alarm. Pressing the ‘ALARM’ button will cause the LED to go steady if any alarms<br />

are still active, or extinguish if there are no active alarms. The horn relay will operate when an alarm occurs.<br />

Pressing ‘ALARM’ will mute the horn relay which will re-initiate on any new alarms.<br />

4.3 ALARM SCHEDULE<br />

4.3.1 SUMMARY OF ALARMS<br />

ALARM DESCRIPTION<br />

‘SENSOR FAIL’ Oxygen cell or electrode failure (high impedance); (inhibited under 650°C).<br />

‘<strong>PROBE</strong> HEATER In the first 20 minutes of power being applied to<br />

FAILURE’ the heater after being switched on, this alarm will not occur, but a ‘<strong>PROBE</strong><br />

UNDER–TEMPERATURE’ display will occur and relay RL8 will be activated.<br />

Refer to Section 6.13.<br />

‘<strong>PROBE</strong> TC O/C’ Probe thermocouple is open circuit. The heater in heated probes will switch off.<br />

‘REF AIR FAIL’ Low reference air flow to probe.<br />

‘FLOW SWITCH’ The reference air flow switch has failed. This is tested by automatically<br />

switching off the reference air pump once every ten minutes for three seconds to<br />

check the operation of the flow switch.<br />

‘A/D CAL ERROR’ The analog to digital converter has been found to fall outside the normal<br />

calibration specifications. This is an electronic fault.<br />

‘D/A CAL ERROR’ The digital to analog and voltage isolator circuit has been found to fall outside<br />

the normal calibration specifications. This check is only performed when the<br />

‘AUTO CAL’ button is pressed. Refer to Section 6.6.<br />