100| Dräger-<strong>Tubes</strong> & CMS-<strong>Handbook</strong><br />

Cross Sensitivity<br />

Dräger-<strong>Tubes</strong> are calibrated for a specific contaminant, but under other than ideal<br />

conditions, other interfering contaminants may also be present. Other contaminants that<br />

influence the indication are described as being cross sensitive.<br />

The information given under the Cross Sensitivity section indicates which contaminants can<br />

influence the indication and which contaminants would not influence the indication.<br />

However, this information does not address all possibilities. The influence of other<br />

contaminants should be reviewed with the Dräger technical services department.<br />

Extension of the Measuring Range<br />

If the standard measuring range of a Dräger-Tube can be expanded by taking more or fewer<br />

pump strokes, the information, including pump strokes, correction factors, etc. will be given.<br />

If there is no information given, please contact the Dräger technical services group.<br />

Additional Information<br />

Supplementary points that must be considered when conducting a Dräger-Tube<br />

measurement are given here.