238| Dräger-<strong>Tubes</strong> for short-term measurements<br />

Phosphine 0.1/c<br />

Order No. 81 03 711<br />

Application Range<br />

Standard Measuring Range: 0.5 to 3 ppm / 0.1 to 1.0 ppm<br />

Number of Strokes n: 1 / 3<br />

Time for Measurement: approx. 1 min / approx. 2.5 min<br />

Standard Deviation: ± 10 to 15 %<br />

Color Change:<br />

yellow ä red<br />

Ambient Operating Conditions<br />

Temperature: 2 to 40 °C<br />

Absolute Humidity:<br />

< 40 mg H 2 O / L<br />

Reaction Principle<br />

HgCI 2 + PH 3 ä Hg-phosphide + HCI<br />

HCI + pH-Indicator ä red reaction product<br />

Cross Sensitivity<br />

A maximum of 6 ppm sulphur dioxide or 15 ppm hydrogen chloride<br />

have no influence on the reasing. Higher concentration result in<br />

plus errors. Ammonia (>100 ppm) result in minus errors. Hyrogen<br />

sulphide and Arsine are indicated with different sensitivity. 30 ppm<br />

hydrocyanic acid does not affect.<br />