310| Measurement of Contaminants in Liquids<br />

Acetic Acid 0.5 to 20 g/L<br />

Order No. 67 22 101<br />

Application Range<br />

Determination of acetic acid in water/waste water<br />

Dräger-Tube: Acetic Acid 5/a<br />

Measuring range:<br />

Number of Strokes (n): 10<br />

Typical Stroke Time:<br />

Measurement Time:<br />

Sample Volume:<br />

Color Change:<br />

0.5 to 20 g/L<br />

10 to 30 s<br />

approx. 200 s<br />

200 mL<br />

blue violet ä yellow<br />

Temperature Range: 10 to 30 °C<br />

pH-Measurement:<br />

necessary<br />

Information of Measurement<br />

Using sulfuric acid, the pH-value has to be adjusted to the value<br />

of 1.3<br />

System Parameters (valid for pH 1.3)<br />

Measurement<br />

Range<br />

[g/L]<br />

Standard<br />

Deviation<br />

[%]<br />

Temperature<br />

[°C]<br />

Parameters<br />

B C<br />

0.5 to 20 25 10 to 30 0.339 1.368<br />

Evaluation of Measurement<br />

Calculate acetic acid concentration:<br />

Y [g/L] = A • B • (X [ppm] + C)<br />

Cross Sensitivity<br />

Formic acid is indicated with lower and propionic acid with higher<br />

sensitivity.<br />

D-13305-2010