Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
310| Measurement of Contaminants in Liquids<br />
Acetic Acid 0.5 to 20 g/L<br />
Order No. 67 22 101<br />
Application Range<br />
Determination of acetic acid in water/waste water<br />
Dräger-Tube: Acetic Acid 5/a<br />
Measuring range:<br />
Number of Strokes (n): 10<br />
Typical Stroke Time:<br />
Measurement Time:<br />
Sample Volume:<br />
Color Change:<br />
0.5 to 20 g/L<br />
10 to 30 s<br />
approx. 200 s<br />
200 mL<br />
blue violet ä yellow<br />
Temperature Range: 10 to 30 °C<br />
pH-Measurement:<br />
necessary<br />
Information of Measurement<br />
Using sulfuric acid, the pH-value has to be adjusted to the value<br />
of 1.3<br />
System Parameters (valid for pH 1.3)<br />
Measurement<br />
Range<br />
[g/L]<br />
Standard<br />
Deviation<br />
[%]<br />
Temperature<br />
[°C]<br />
Parameters<br />
B C<br />
0.5 to 20 25 10 to 30 0.339 1.368<br />
Evaluation of Measurement<br />
Calculate acetic acid concentration:<br />
Y [g/L] = A • B • (X [ppm] + C)<br />
Cross Sensitivity<br />
Formic acid is indicated with lower and propionic acid with higher<br />
sensitivity.<br />
D-13305-2010