Application Range<br />

Determination of formic acid in water/waste water<br />

Dräger-Tube: Acetic Acid 5/a<br />

Measuring range:<br />

Number of Strokes (n): 10<br />

Typical Stroke Time:<br />

Measurement Time:<br />

Sample Volume:<br />

Color Change:<br />

Formic Acid 1 to 20 g/L<br />

Order No. 67 22 101<br />

1 to 20 g/L<br />

10 to 30 s<br />

approx. 200 s<br />

200 mL<br />

blue violet ä yellow<br />

Temperature Range: 5 to 25 °C<br />

pH-Measurement:<br />

necessary<br />

|331<br />

Information of Measurement<br />

Using sulfuric acid, the pH-value has to be<br />

adjusted to the value of 1.3.<br />

System Parameters (valid for pH 1.3)<br />

Measurement<br />

Range<br />

[g/L]<br />

Standard<br />

Deviation<br />

[%]<br />

Temperature<br />

[°C]<br />

Parameters<br />

B C<br />

1 to 20 25 5 to 25 0.487 1.607<br />

E<br />

Evaluation of Measurement<br />

Calculate formic acid concentration:<br />

Y [g/L] = A • B • (X [ppm] + C)<br />

Cross Sensitivity<br />

Acetic acid and propionic acid are indicated with higher<br />

sensitivity.<br />