66| Dräger-<strong>Tubes</strong> & CMS-<strong>Handbook</strong><br />

3.2 The Components of the Dräger CMS<br />

The Chip-Measurement-System is a system for the<br />

quant itative determination of hazardous gas or vapor<br />

concentrations in air. The measurements are carried out in<br />

the work place to monitor for hazardous gas concentrations,<br />

process control, and for measurements in confined spaces,<br />

etc. This system is designed for short-term measurements.<br />

The complete measuring system consists of two main key<br />

components:<br />

ST-14162-2008<br />

- Substance-Specific Chips<br />

- Analyzer<br />

ST-1347-2004<br />

The Chip-Measurement-System<br />

The Chip-Measurement-System combines the advantages of five components:<br />

- the chip as capillary array for 10 possible measurements,<br />

- optics for detection of the reaction products,<br />

- the pump system with mass-flow controller to draw in ambient air and ensure a constant<br />

air mass-flow,<br />

- mechanics designed to position the chip into the analyzer and to open and test the<br />

corresponding capillary in preparation for measuring,<br />

- electronics and software to control and regulate the measuring procedure, to process<br />

the signal and of course to provide a digital display of the measured concentrations.