18.05.2018 Views

Tubes Handbook 18th Edition (EN) - May 2018

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Carbon Dioxide 100/a-P<br />

Order No. 67 28 521<br />

|297<br />

Application Range<br />

Use in Aerotest 5000, Aerotest Alpha, MultiTest med. Int.,<br />

Aerotest HP<br />

Standard Measuring Range: 100 to 3,000 ppm<br />

Test Volume:<br />

1 L<br />

Flow Rate:<br />

0,2 L / min<br />

Time for Measurement:<br />

approx. 5 min<br />

Standard Deviation: ± 10 to 15 %<br />

Color Change:<br />

white ä violet<br />

A<br />

Ambient Operating Conditions<br />

Temperature: 15 to 25 °C<br />

Absolute Humidity:<br />

max. 23 mg H 2 O / L<br />

Pressure:<br />

The tube may only be<br />

used for depressurized<br />

compressed air<br />

Reaction Principle<br />

CO 2 + N 2 H 4 ä NH 2 -NH-COOH Crystal violet<br />

Cross Sensitivity<br />

Hydrogen sulfide and sulfur dioxide in the TLV range are not<br />

indicated.<br />

ST-51-2001