18.05.2018 Views

Tubes Handbook 18th Edition (EN) - May 2018

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

120| Dräger-<strong>Tubes</strong> for short-term measurements<br />

Benzene 5/a<br />

Order No. 67 18 801<br />

Application Range<br />

Standard Measuring Range: 5 to 40 ppm<br />

Number of Strokes n: 15 to 2<br />

Time for Measurement:<br />

max. 3 min<br />

Standard Deviation: ± 30 %<br />

Color Change:<br />

white ä red brown<br />

Ambient Operating Conditions<br />

Temperature: 0 to 40 °C<br />

Absolute Humidity:<br />

max. 50 mg H 2 O / L<br />

Reaction Principle<br />

2 C 6 H 6 + HCHO ä C 6 H 5 -CH 2 -C 6 H 5 + H 2 O<br />

C 6 H 5 -CH 2 -C 6 H 5 + H 2 SO 4 ä p-quinoid compound<br />

Cross Sensitivity<br />

Other aromatics (toluene, xylene) are retained in the pre-layer<br />

causing a reddish brown discoloration. If the toluene or xylene<br />

concentrations are too high the entire pre-layer up to the indicating<br />

layer is discolored making a benzene measurement impossible. Petroleum<br />

hydrocarbons, alcohols and esters do not affect the<br />

indication.<br />

ST-22-2001

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!