334| Measurement of Contaminants in Liquids<br />

Hydrogen Sulfide 50 to 500 µg/L<br />

Order No. 81 01 461<br />

Application Range<br />

Determination of hydrogen sulfide (sum of sulfide) in water/waste<br />

water<br />

Dräger-Tube:<br />

Hydrogen Sulfide 0.2/a<br />

Measuring range:<br />

50 to 500 mg/L<br />

Number of Strokes (n): 5<br />

Typical Stroke Time:<br />

50 to 80 s<br />

Measurement Time:<br />

approx. 325 s<br />

Sample Volume:<br />

200 mL<br />

Color Change:<br />

white ä pale brown<br />

Temperature Range: 3 to 30 °C<br />

pH-Measurement:<br />

necessary<br />

Information of Measurement<br />

Using acetic acid or sodium hydroxide solution, the pH-value has to<br />

be adjusted to the value of 7.3 - 7.4 (K=1)<br />

System Parameters (valid for pH 7.3 - 7.4)<br />

Measurement<br />

Range<br />

[µg/L]<br />

Standard<br />

Deviation<br />

[%]<br />

Temperature<br />

[°C]<br />

Parameters<br />

B C<br />

50 to 500 30 3 to 7 72 0.2<br />

8 to 13 63 0.2<br />

14 to 30 57 0.2<br />

Evaluation of Measurement<br />

Calculate hydrogen sulfide concentration:<br />

Y [µg/L] = A • B • (K •X [ppm] + C)<br />

ST-132-2001