Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
74| Dräger-<strong>Tubes</strong> & CMS-<strong>Handbook</strong><br />
The basis of this validation was the Austrian standard <strong>EN</strong> 482:<br />
"Workplace Atmospheres –<br />
General Requirements for Methods of Measuring Chemical Agents".<br />
The institute confirms:<br />
- The measuring results of the Dräger CMS correspond with the results of<br />
the reference methods.<br />
- The accuracy of the Dräger CMS is far higher than the required accuracy of<br />
the standard <strong>EN</strong> 482.<br />
- The Dräger CMS is a suitable measurement system.<br />
3.9 Performance Data of the Dräger CMS<br />
Measuring Range<br />
depends on the chip type<br />
Typical Measuring Time 20 seconds to 3 minutes (depends on the chip type and<br />
the concentration of hazardous gas to be measured)<br />
20 seconds to 10 minutes with special hazardous gas<br />
Calibration<br />
by manufacturer<br />
Temperature (operation) 0 to 40 °C (32 to 104 °F)<br />
Temperature (storage)<br />
- 20 to 60 °C (- 4 to 140 °F) (Analyzer)<br />
< 25 °C (< 77 °F) (Chips)<br />
Atmospheric Pressure<br />
700 to 1,100 hPa<br />
Relative Humidity<br />
0 to 95 %, not condensing<br />
System Diagnosis<br />
automatic self-diagnosis, microprocessor controlled<br />
for all components of the system<br />
Display<br />
LCD, alphanumerical digits, backlight<br />
Menu Languages<br />
English, German, French, Spanish<br />
Operational Time<br />
approx. 450 minutes of measurements<br />
(per battery pack)<br />
Power Supply<br />
4 x 1.5-V-batteries, only use the following types:<br />
Ralsten (Energizer) Alkaline LR6<br />
Duracell MN 1500 LR6<br />
Rayovak Rechargeable Alkaline AA<br />
(only to use with Rayovak Charger PS1 or PS3)<br />
Weight<br />
730 g (1.6 lbs., analyzer with batteries)<br />
Dimensions 215 x 105 x 65 mm; 8.5" x 4.1" x 2.5" (L x W x H)<br />
Electromagnetic Interference EMV <strong>EN</strong> 550 11; by 10/01 <strong>EN</strong> 502 70<br />
Radio Frequencey Interference RFI <strong>EN</strong> 550 14<br />
Ingress Protection IP 54