Chapter 2 Specifications<br />

Concentration Alarm Low, High, and STEL<br />

Power Requirements of<br />

ac/dc Converter<br />

Measurement Cycle Time<br />

Reference, Operating, and<br />

Storage Conditions<br />

Electromagnetic<br />

Compatibility (EMC)<br />

95 to 260 V ac at 50/60 Hz, ±3% or 11 to 15 V dc, 5 A<br />

(from automobile battery with optional dc power cable)<br />

Single gas analysis screen updated each half-second.<br />

Dependent upon application in multigas mode or if<br />

compensations are enabled.<br />

See Table 2-2.<br />

Complies with the requirements of the European EMC<br />

Directive 89/336/EEC.<br />

Sample Flow Rate<br />

14 liters/minute (nominal) with sampling wand and<br />

particulate filter.<br />

Sample Cell Volume 2.23 liters<br />

Gas Purging 5 volumes of sample required to purge cell by 99%.<br />

Table Error! No text of specified style in document.-2. Operating and Storage<br />

Conditions<br />

Influence<br />

Ambient<br />

Temperature<br />

Relative<br />

Humidity<br />

Ambient<br />

Pressure (d)<br />

Reference<br />

Operating<br />

Conditions (a)<br />

23 ±2°C<br />

(73 ±4°F)<br />

50 ±10%<br />

101.3 ±1% kPa<br />

(14.7 ±1% psi)<br />

Normal<br />

Operating<br />

Conditions (b)<br />

5 to 40°C<br />

(40 to 105°F)<br />

5 to 95%<br />

(noncondensing)<br />

86 to 106 kPa<br />

(12.5 to 15.3 psi)<br />

2-2<br />

Normal<br />

Operative<br />

Limits (c)<br />

1 and 50°C<br />

(35 and 120°F)<br />

0 and 95%<br />

(noncondensing)<br />

80 and 110 kPa<br />

(11.6 and 15.9<br />

psi)<br />

Transportation<br />

and<br />

Storage Limits<br />

(c)<br />

-20 and +60°C<br />

(-5 and 140°F)<br />

0 and 100%<br />

(noncondensing)<br />

80 and 110 kPa<br />

(11.6 and 15.9<br />

psi)<br />

(a) Range of operating conditions within which the influence is negligible.<br />

(b) Range of operating conditions within which a device is designed to operate and for which the<br />

specifications for the influence applies.<br />

(c) Range to which a device can be subjected without permanent impairment of operating<br />

characteristics. Device may require adjustment to restore specified performance upon returning<br />

within the limits of Normal Operating Conditions. Maximum temperature for intrinsically safe<br />

models is 40°C<br />

(d) Single pressure point within the range, pressure may not vary throughout the range.