Chapter 2 Specifications<br />
Concentration Alarm Low, High, and STEL<br />
Power Requirements of<br />
ac/dc Converter<br />
Measurement Cycle Time<br />
Reference, Operating, and<br />
Storage Conditions<br />
Electromagnetic<br />
Compatibility (EMC)<br />
95 to 260 V ac at 50/60 Hz, ±3% or 11 to 15 V dc, 5 A<br />
(from automobile battery with optional dc power cable)<br />
Single gas analysis screen updated each half-second.<br />
Dependent upon application in multigas mode or if<br />
compensations are enabled.<br />
See Table 2-2.<br />
Complies with the requirements of the European EMC<br />
Directive 89/336/EEC.<br />
Sample Flow Rate<br />
14 liters/minute (nominal) with sampling wand and<br />
particulate filter.<br />
Sample Cell Volume 2.23 liters<br />
Gas Purging 5 volumes of sample required to purge cell by 99%.<br />
Table Error! No text of specified style in document.-2. Operating and Storage<br />
Conditions<br />
Influence<br />
Ambient<br />
Temperature<br />
Relative<br />
Humidity<br />
Ambient<br />
Pressure (d)<br />
Reference<br />
Operating<br />
Conditions (a)<br />
23 ±2°C<br />
(73 ±4°F)<br />
50 ±10%<br />
101.3 ±1% kPa<br />
(14.7 ±1% psi)<br />
Normal<br />
Operating<br />
Conditions (b)<br />
5 to 40°C<br />
(40 to 105°F)<br />
5 to 95%<br />
(noncondensing)<br />
86 to 106 kPa<br />
(12.5 to 15.3 psi)<br />
2-2<br />
Normal<br />
Operative<br />
Limits (c)<br />
1 and 50°C<br />
(35 and 120°F)<br />
0 and 95%<br />
(noncondensing)<br />
80 and 110 kPa<br />
(11.6 and 15.9<br />
psi)<br />
Transportation<br />
and<br />
Storage Limits<br />
(c)<br />
-20 and +60°C<br />
(-5 and 140°F)<br />
0 and 100%<br />
(noncondensing)<br />
80 and 110 kPa<br />
(11.6 and 15.9<br />
psi)<br />
(a) Range of operating conditions within which the influence is negligible.<br />
(b) Range of operating conditions within which a device is designed to operate and for which the<br />
specifications for the influence applies.<br />
(c) Range to which a device can be subjected without permanent impairment of operating<br />
characteristics. Device may require adjustment to restore specified performance upon returning<br />
within the limits of Normal Operating Conditions. Maximum temperature for intrinsically safe<br />
models is 40°C<br />
(d) Single pressure point within the range, pressure may not vary throughout the range.