3-7<br />

Chapter 3 Instrument Preparation<br />

Figure 3-7. Rear of Charger/Power Supply (Model 5100A Shown)<br />

Battery Indicator Calibration<br />

The purpose of the battery indicator calibration is to recalibrate the battery's internal<br />

computer chip. Since the battery's full run time is constantly diminishing by small<br />

amounts, the baseline must be constantly adjusted. This calibration procedure<br />

compensates for this change and recalibrates the system to ensure that the battery<br />

indicator presents as accurate a run-time as possible. A calibration should be performed:<br />

• If there is a prompt to calibrate. This occurs after the analyzer has undergone<br />

64 charge cycles without discharging and fully charging the battery.<br />

• If the analyzer has gone unused for a period of 30 or more days.<br />

Note: Not performing this calibration does not invalidate any of the concentration data.<br />

Warning: Intrinsically safe models require special actions with respect to the following<br />

section. Refer to Appendix B for descriptions of these actions.<br />

The calibration procedure is:<br />

1. Remove the battery from the analyzer and connect it to the Charger/Power<br />

Supply using the battery pack interface cable. See Figure 4-3 for location of<br />

battery compartment. Open compartment using special key (Part Number<br />

X0173QH). Remove battery by partially turning each screw until the battery is<br />

loose.