Chapter 4 Operation<br />

While the cell fills with zero air, the Zero Purge screen indicates the time remaining.<br />

Press ENTER to shortcut the purge.<br />

CAUTION: The zero process requires five volumes of air (about 11 liters) to completely<br />

flush the cell. Shortcutting this step will result in an incomplete purge and incorrect<br />

readings during analysis due to an invalid baseline value.<br />

┌────────────────────────────────────────┐<br />

│╒══CURRENT APPLICATION══╕Wait for cell │<br />

││ │ to fill with │<br />

││ │ zero air...│<br />

││ │ │<br />

││ METHANE: ---- ppm │ 15 seconds │<br />

││ │ │<br />

││ │ │<br />

│└───────────────────────┘Enter=SkipPurge│<br />

└────────────────────────────────────────┘<br />

Zero Purge Display<br />

The instrument automatically begins the zeroing process. Wait while the zero<br />

measurement is taken. The screen displays the percent done. Use the up and down arrow<br />

keys to toggle the measurement views.<br />

┌────────────────────────────────────────┐<br />

│▐██CURRENT APPLICATION██▌ │<br />

││ │ │<br />

││ │ Wait while │<br />

││ │ zeroing...│<br />

││ METHANE: ---- ppm │ [ ] │<br />

││ │ 50% │<br />

││ │ │<br />

│└───────────────────────┘ Esc=Cancel │<br />

└────────────────────────────────────────┘<br />

Zeroing Process Display<br />

4-8