Chapter 4 Operation<br />

CHANGING THE PATHLENGTH<br />

The SapphIRe analyzer can be set to one of two pathlengths: short (0.5 m) and long (12.5<br />

m). To change the pathlength from the short pathlength (0.5 m) to the long pathlength<br />

(12.5 m), turn the hex recess adjustment clockwise with the pathlength change tool or<br />

9/64 hex wrench. The location of the pathlength adjustment is shown in Figure 4-3. To<br />

change the pathlength from long to short, turn the adjustment counterclockwise. See<br />

Location of Pathlength Adjustment for location of adjustment.<br />

Caution: Do not overturn adjustment.<br />

The rotation angle is approximately 90 degrees but the actual short path position<br />

corresponds to the point at which a push pin engages a notch in a cam. As the pin and<br />

cam are aligned, an indicator rod engages a microswitch which tells the microprocessor<br />

that the gas cell is now at short path.<br />

As the mechanism is being changed from the long to short position, there is a slight<br />

increase in resistance to turning the adjustment. This is normal (caused by the<br />

compression of an internal spring) and should not be construed as a defect. An audible<br />

"click" indicates when the mirror is in the proper position.<br />

Short<br />

Long<br />

Figure 4-3. Location of Pathlength Adjustment and Battery Compartment<br />

4-16<br />

Battery<br />

compartment