Chapter 4 Operation<br />
Table 4-4. Warning Messages<br />
Warning Message Meaning Corrective Action<br />
Absorbance is too negative!<br />
Try rezeroing the analyzer<br />
and make sure zero gas is<br />
not contaminated.<br />
Analyzer factory<br />
configuration is invalid!<br />
Analyzer memory is<br />
invalid.<br />
Battery pack capacity<br />
getting low!<br />
Battery pack charge is low!<br />
Not enough power to run<br />
analyzer.<br />
Battery pack data is invalid!<br />
Battery pack needs full<br />
charge, then discharge to<br />
update capacity data.<br />
This message is generated if the measured<br />
absorbance falls below a negative<br />
absorbance threshold value set in the<br />
application. Try rezeroing the analyzer to<br />
correct the problem<br />
4-14<br />
Rezero the analyzer. If<br />
problem persists,<br />
replace the zero gas<br />
filter.<br />
This message is displayed only during the<br />
power-on self-test if the factory<br />
Return analyzer to TEI<br />
configuration is not valid. The analyzer can for service.<br />
operate with a minimum feature set.<br />
The data in the analyzer's battery-backed up<br />
RAM is invalid. All settings are reset to<br />
factory default values if this message is<br />
displayed. If the analyzer power is cycled<br />
again and this message persists, the backup<br />
battery may be dead.<br />
This message indicate the battery pack has<br />
reached the end of its lifetime.<br />
This message indicates the battery is too<br />
low to run the system. The system is shut<br />
down shortly after this message is<br />
displayed.<br />
This message indicates the battery pack has<br />
not been initialized.<br />
This message is displayed if the battery has<br />
not been fully discharge before it has been<br />
recharged many times.<br />
Contact TEI for<br />
assistance.<br />
Replace the battery<br />
pack with a new one.<br />
To continue operation,<br />
replace the battery with<br />
a fresh battery or<br />
connect the analyzer to<br />
a charger.<br />
Contact TEI customer<br />
service for a new<br />
battery pack.<br />
Perform battery<br />
indicator calibration<br />
(see page 3-7).