01.02.2013 Views

Thermo Miran Sapphire XL Manual 2005 - IE Monitoring Instruments

Thermo Miran Sapphire XL Manual 2005 - IE Monitoring Instruments

Thermo Miran Sapphire XL Manual 2005 - IE Monitoring Instruments

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Chapter 5<br />

ANALYZE/START<br />

To start the analysis, press 1 (Analyze/Start) from the Main Menu. The display presents<br />

the Zero Menu if a zero has been taken for all of the current application wavelengths.<br />

The date and time of the last zero and analyzer status are shown. To re-zero the analyzer,<br />

press 1. To continue with an analysis without taking a new zero value, press ENTER.<br />

Note: If the analyzer has not been recently zeroed, the analyzer skips the Zero Menu<br />

and displays the Analysis menu. See “To Begin Analysis” later in this chapter.<br />

TO ZERO THE ANALYZER<br />

┌────────────────────────────────────────┐<br />

│╒══CURRENT APPLICATION══╕ ZERO MENU │<br />

││ │Last zero: │<br />

││ │ 04Jan01 13:30 │<br />

││ │ │<br />

││ METHANE: ---- ppm │Status: Valid │<br />

││ │ │<br />

││ │ 1=New zero │<br />

│└───────────────────────┘Enter=Continue │<br />

└────────────────────────────────────────┘<br />

Zero Menu<br />

Caution:<br />

1. When zeroing the analyzer, it is important to zero the unit in the area where the<br />

analysis is to take place. This assures that the analyzer is zeroed with air (scrubbed by<br />

the zero gas filter) at the same temperature and humidity as the sample to be analyzed.<br />

2. However, for chemicals such as Butane, Carbon Dioxide, Carbon Monoxide, Ethane,<br />

Methane, Nitrous Oxide, Propane, and Water Vapor, that can not be scrubbed by the zero<br />

gas filter, it is important zero the unit in an area free of the chemical. For example, if<br />

methane is being measured, zero the analyzer in a Methane free area. (Note: typical<br />

atmospheric conditions are 2-3 ppm.) If the analyzer is zeroed in an area containing 7<br />

ppm of Methane, the zero value is set at 7 ppm. All subsequent readings are then<br />

referenced to the 7 ppm baseline value. Thus, if Methane rises to 15 ppm, the analyzer<br />

displays a reading of 8 ppm; if it drops to 3 ppm, the display reads -4 ppm.<br />

3. TEI recommends that the zero gas filter be replaced after 30 zero cycles.<br />

4. The purging of the gas cell during the zero process requires five volumes of air (about<br />

11 liters). The timing supplied in the analyzer's zero routine is ample to allow for this to<br />

happen. However, if there is a restriction in flow that reduces the sample flow rate below<br />

approximately 11 liters/minute or if the sample hose is lengthened, it is possible that the<br />

cell may not be satisfactorily purged under these conditions. If so, repeat the zero<br />

process until sufficient sample has been drawn through the zero filter.<br />

5-1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!