ERROR/FAULT MESSAGES<br />

5-7<br />

Chapter 5 Analysis<br />

The instrument displays error/fault messages relating to analysis by indicators on the<br />

display or by flag screens.<br />

Display Indicators<br />

A letter or symbol at the end of the analyzer reading indicates the following:<br />

Table 4-2. Display Indicators<br />

Symbol Description<br />

H High alarm level exceeded<br />

L Low alarm level exceeded<br />

S Short Term Exposure Limit alarm level exceeded<br />

X Fault/error<br />

P Purge not done<br />

Z Zero invalid<br />

> Concentration exceeds full scale value (HRL).<br />

Measurements are below the range of the measurement electronics. Possible<br />

�<br />

corrective action is to rezero the analyzer or change the pathlength.<br />

Measurements exceed the range of the measurement electronics. Possible<br />

�<br />

corrective action is to rezero the analyzer or change the pathlength.<br />

Calculation overflow where range has been exceeded for processing a<br />

}<br />

measurement value. Possible corrective action is to repeat the measurement.