40 User Guide <strong>OSMOMAT</strong> <strong>070</strong> Version 1.1<br />

Figure 4-12: Example: Result: Molecular weight<br />

Linear regression is performed automatically. The intercept is the result that indicates how great the<br />

measurement effect is for a substance concentration of 1g/kg of solvent.<br />

Figure 4-13: Example: Linear regression<br />

To calculate the molecular weight, the results of the calibration measurement [(measurement<br />

effect)/(mol/kg)] are compared with the result of the sample measurement [(measurement<br />

effect)/(g/kg)]. The final result is the number average of the molecular weight Mn determined by the<br />

program.<br />