6 User Manual <strong>OSMOMAT</strong> <strong>070</strong> Version 1.2<br />
The following items are located on the rear of the unit:<br />
Figure 1-2 Osmomat <strong>070</strong>, Rear<br />
8<br />
(5) Socket for connecting the cell head<br />
(6) Potentiometer for fine adjustment of the head temperature<br />
(7) Connector sockets A and B for the measuring probes<br />
(8) Power connector for plugging the cell unit into the mains voltage via the control unit<br />
(9) Mains fuse for two-phase protection of the cell unit 230V/2.5A, 115V/4A<br />
(10) Nameplate with information on the unit type, serial number, supply voltage, power consumption<br />
and line frequency<br />
(11) Connector socket for connection to the control-unit<br />
(12) Connector socket for connection of a recorder<br />
1.2.3 Design of the measurement cell<br />
The beaker-shaped measurement cell is located in a thermostat with a very constant temperature that<br />
can be adjusted in one-degree stages between 30°C and 130°C. The cell thermostat surrounds a<br />
cylindrical beaker that contains a small amount of solvent (filling level is about 1 cm). To achieve 100%<br />
vapor saturation, the beaker contains vapor wicks in the form of filter paper strips.<br />
Two measuring probes with temperature sensors are suspended in the cell space saturated with solvent<br />
vapor. The bottom ends of the measuring probes are coated with drops of solvent or solution. An even<br />
drop size is achieved through wire coils attached to the bottom end of the measuring probes. They make<br />
it easier to rinse the probes, because the liquid can drain off spirally and the probes can be thoroughly<br />
wetted. The solvents or solutions are fed through glass syringes that are placed in the cell head along<br />
with the measuring probes.<br />
9<br />
11<br />
10 12<br />
6 User Manual <strong>OSMOMAT</strong> <strong>070</strong> Version 1.2<br />
5<br />
6<br />
7