8 User Guide <strong>OSMOMAT</strong> <strong>070</strong> Version 1.1<br />
There are two measuring channels in the cell and cell head for external temperature checking.<br />
1.2.4 The cell head<br />
The cell head has its own thermostat that automatically adjusts itself to the cell temperature. It closes<br />
the measurement cell upward and ensures that the syringes are preheated. Fine adjustment of the<br />
temperature to optimize the temperature difference between the cell and cell head can be carried out<br />
with a potentiometer on the rear Figure 1-2 (6). The temperature of the cell head should be about<br />
0.5°C lower than the cell temperature. Temperatures can be checked with an external thermometer in<br />
the corresponding measuring channels Figure 1-6: Measurement of the head/cell temperature. To<br />
prevent temperature effects, the cell head has a cover that is fastened with a knurled screw.<br />
Figure 1-4: Osmomat <strong>070</strong> cell head<br />
29-Teflon lip seal for sealing the cell space<br />
30-Guide tube for aligning the syringe needles to the dropwise addition point of the measuring probes<br />
8 User Guide <strong>OSMOMAT</strong> <strong>070</strong> Version 1.1