If the KVO rate is set to 0, the pump will beep ten minutes prior to the beginning of the<br />

Continuous Therapy start time. To stop the audible beep, press the “Silence” key.<br />

IMPORTANT<br />

INFORMATION<br />

During the delay start period, if the KVO rate is set to 0, no fluid is<br />

infusing. Be sure the patient’s access device is patent and has been<br />

flushed per facility policy to maintain patency during this time period.<br />

If a KVO rate is programmed, then that KVO rate will be used in the<br />

Delay Start period.<br />

4 TITRATE: Allows the programmer to set titration limits into the prescription settings of the<br />

pump. If there is no titration prescription, select “OFF” and press “YES/ENTER” to<br />

accept. The cursor will move to the next field. If a titration prescription has been written,<br />

press the “NO” key to select “ON” and press the “YES/ENTER” key to accept. The<br />

following screen will appear to program the titration parameters:<br />

Continuous Titration Limit Screen<br />

a. MaxRATE (Maximum Rate) limits: Use the numeric keys to enter the upper limits of the<br />

infusion rate. When the maximum titration rate is entered correctly, press<br />

“YES/ENTER” and the cursor moves to the next field. The range for this field is from<br />

0.1 to 400 ml/hr.<br />

b. ACCEPT?: If there are any changes to be made, use the up or down arrows to move to<br />

the desired field, use the numeric keys to enter the corrected data, and press<br />

“YES/ENTER” to accept the new values. Move the cursor back to the “ACCEPT?”<br />

field and enter “YES/ENTER”. The Titration Limit is now entered, and the display<br />

returns to the Pre-Prescription Screen. If the entire Titration Limit Screen is to be<br />

aborted, press “NO” to return to the Pre-Prescription Screen.<br />

5 NEXT?: If there are any changes to be made, use the up or down arrow keys to locate the<br />

field to be changed, use the numeric keys to enter the corrected information, and press<br />

“YES/ENTER” to accept the new data. If all the fields are acceptable, move the cursor to<br />

the “NEXT?” field and press the “YES/ENTER” key to move on to the next screen.<br />

At the following screen, the prescription information is entered to complete the programming of<br />

the pump for a Continuous Therapy. Note that the Screen Identifier, at the left side of the display,<br />

gives the abbreviation “CON” for Continuous Therapy and has a down arrow configuration,<br />

which indicates that there are more than four fields of text or lines of information to be<br />

considered.<br />

Continuous Therapy<br />

43