<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Introduction<br />
Introduction<br />
The <strong>CETAC</strong> <strong>Marin</strong>-5 Enhanced Nebulizer System is designed primarily<br />
for ICP-AES and ICP-MS. The <strong>Marin</strong>-5 consists of a pneumatic<br />
nebulizer, a specialized heated spray chamber, and a dedicated<br />
desolvation apparatus.<br />
Sensitivity is improved with the <strong>Marin</strong>-5 by increasing analyte<br />
transport efficiency and reducing solvent loading to the plasma.<br />
Functionally, liquid sample is pumped or self-aspirated to a pneumatic<br />
nebulizer at an uptake rate of ≤ 0.5 mL/min and an argon nebulizer gas<br />
flow rate of 0.5 to 1.1 L/min. The nebulizer is inserted into a heated<br />
spray chamber to convert a higher percentage of the liquid into an<br />
aerosol. The spray chamber is connected to a desolvation system to<br />
condense most of the liquid in the aerosol and maintain plasma<br />
stability. A built-in three-channel drain pump removes the condensed<br />
sample solvent to waste. Dried sample particles then continue in the<br />
nebulizer gas flow to the plasma, resulting in emitted analyte radiation<br />
(ICP-AES) or analyte ions (ICP-MS).<br />
The nebulizer, spray chamber, and desolvation system are protected by<br />
a secure lid, with access to the nebulizer via a front door. An external<br />
view of the <strong>Marin</strong>-5 is shown in Figure 1-1.<br />
1–2