<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />

Maintaining the <strong>Marin</strong>-5<br />

13 Reestablish power to the <strong>Marin</strong>-5.<br />

__________________________________________________________________<br />

Rinse Procedure for the <strong>Marin</strong>-5 Condenser<br />

The <strong>Marin</strong>-5 condenser section may also be rinsed using the same<br />

solution guidelines as for the <strong>Marin</strong>-5 glassware. To rinse the condenser<br />

section, perform the following steps:<br />

1 Turn the AC power switch to off (0) and disconnect the AC<br />

power cord. If the power for the <strong>Marin</strong>-5 has been on, allow the<br />

unit to cool for 30 minutes.<br />

2 Remove the sample out line from the glass ball joint on the back<br />

of the <strong>Marin</strong>-5.<br />

3 Use a rinse bottle to add 20mL to 25mL of the appropriate rinse<br />

solution to the condenser.<br />

4 Allow the rinse solution to stand in the condenser for 10 to 15<br />

minutes.<br />

5 Reattach the power cord to the <strong>Marin</strong>-5 and turn on the power.<br />

The drain pump will automatically turn on and remove the<br />

rinse solution to waste. When all of the rinse solution has been<br />

removed, turn off the power and disconnect the power cord.<br />

6 Steps 3 through 5 may be repeated with deionized water.<br />

7 Reestablish power to the <strong>Marin</strong>-5 and reconnect the sample out<br />

line.<br />

6-8