Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Maintaining the <strong>Marin</strong>-5<br />
13 Reestablish power to the <strong>Marin</strong>-5.<br />
__________________________________________________________________<br />
Rinse Procedure for the <strong>Marin</strong>-5 Condenser<br />
The <strong>Marin</strong>-5 condenser section may also be rinsed using the same<br />
solution guidelines as for the <strong>Marin</strong>-5 glassware. To rinse the condenser<br />
section, perform the following steps:<br />
1 Turn the AC power switch to off (0) and disconnect the AC<br />
power cord. If the power for the <strong>Marin</strong>-5 has been on, allow the<br />
unit to cool for 30 minutes.<br />
2 Remove the sample out line from the glass ball joint on the back<br />
of the <strong>Marin</strong>-5.<br />
3 Use a rinse bottle to add 20mL to 25mL of the appropriate rinse<br />
solution to the condenser.<br />
4 Allow the rinse solution to stand in the condenser for 10 to 15<br />
minutes.<br />
5 Reattach the power cord to the <strong>Marin</strong>-5 and turn on the power.<br />
The drain pump will automatically turn on and remove the<br />
rinse solution to waste. When all of the rinse solution has been<br />
removed, turn off the power and disconnect the power cord.<br />
6 Steps 3 through 5 may be repeated with deionized water.<br />
7 Reestablish power to the <strong>Marin</strong>-5 and reconnect the sample out<br />
line.<br />
6-8