Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Installing the <strong>Marin</strong>-5<br />
4mm O.D. tubing or may be equipped with an adapter,<br />
depending on the ICP-AES or ICP-MS instrument.<br />
15 Open the front door of the <strong>Marin</strong>-5 cover. Insert the other end<br />
of the nebulizer gas line through the sample out port on the<br />
back of the <strong>Marin</strong>-5 and into the interior of the <strong>Marin</strong>-5.<br />
Connecting the Nebulizer<br />
The <strong>Marin</strong>-5 is provided with one (1) concentric nebulizer. To install the<br />
nebulizer and the nebulizer gas line, follow the steps below:<br />
1 Carefully insert the nebulizer into the nebulizer adapter that<br />
is mounted in the spray chamber. Note that the adapter hole<br />
will be angled at 5 or 15 degrees, so insert the nebulizer slowly.<br />
2 Connect the open end of the 4mm tubing (of the nebulizer gas<br />
line) to the press-fit adapter mounted on the side gas port of<br />
the nebulizer. See Figure 3- 9.<br />
Figure 3- 9. Connection of Nebulizer Gas Line<br />
3–10