Oxygen addition may be needed to prevent carbon buildup from organic<br />

solvents on the ICP torch or ICP-MS sampling cones (sampler and<br />

skimmer). Argon or nitrogen addition gas can help improve analyte<br />

signal intensity and/or stability, expecially for ICP-MS. Flow rates for<br />

these gases can vary considerably depending upon the application and<br />

the model ICP or ICP-MS. Typical flow rate ranges are:<br />

a O2 : 10 to 50mL/min<br />

b Ar : 0.05 to 0.20 L/min<br />

c N2 : 3 to 10 mL/min<br />

A picture of the tee assembly is given in Figure 4- 3.<br />

Figure 4- 3. Tee Assembly for Sample Out Line<br />

4-8