03.01.2015 Views

Marin-5 Operator's Manual - CETAC Technologies

Marin-5 Operator's Manual - CETAC Technologies

Marin-5 Operator's Manual - CETAC Technologies

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />

Troubleshooting the <strong>Marin</strong>-5<br />

• Low or No signal from the ICP-AES or ICP-MS. Check the<br />

sample out line to make certain the Tygon® tubing is properly<br />

connected to the torch.<br />

• Check the o-rings in the nebulizer port on the front of the<br />

spray chamber. Worn-out o-rings may cause gas to leak out of the<br />

spray chamber. Turn off power to the <strong>Marin</strong>-5 and allow the spray<br />

chamber to cool (30 minutes). Then remove the nebulizer port fitting<br />

and replace the o-rings. Note that there are 2 sets of o-rings (outer<br />

and inner). See Figure 7-1.<br />

Figure 7-1. Nebulizer Port with o-rings.<br />

2 Sample Flow Rate Problems<br />

• The liquid flow rate is too high. The liquid flow rate by selfaspiration<br />

is determined by the nebulizer gas flow. Decreasing the<br />

nebulizer gas flow will decrease the liquid flow rate. Consequently,<br />

signal intensity may decrease as a result. Proper nebulizer gas flow<br />

rates should be optimized based on each individual application.<br />

Normal criterion for assessing performance are signal to<br />

7-4

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!