<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Installing the <strong>Marin</strong>-5<br />
Connecting the <strong>Marin</strong>-5 Drain Tubing<br />
There are three liquid drain lines to install on the back of the <strong>Marin</strong>-5<br />
unit. To set up these drain lines:<br />
1. Attach the individual (3 each) drain lines to the barbed<br />
fittings on the bottom of the drain peristaltic pump. See<br />
Figure 3- 12.<br />
Figure 3- 12. Attachment of Drain Tubing<br />
2. The outlet of the drain lines may be placed in a suitable<br />
waste container.<br />
NOTE:<br />
The drain tubing supplied is Teflon® FEP-lined Tygon® tubing. When<br />
attaching this tubing, be careful that the inner lining does not come<br />
away and block the flow of waste liquid. If the lining does come away,<br />
the affected area may be removed and the remaining length of tubing<br />
attached.<br />