03.01.2015 Views

Marin-5 Operator's Manual - CETAC Technologies

Marin-5 Operator's Manual - CETAC Technologies

Marin-5 Operator's Manual - CETAC Technologies

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />

Maintaining the <strong>Marin</strong>-5<br />

__________________________________________________________________<br />

Replacement of the Drain Pump Tubing<br />

The drain peristaltic pump tubing may need to be replaced due to wear<br />

or damage. If liquid is not being drained from the glassware, first adjust<br />

the tension on the peristaltic pump tubing. If any adjustment does not<br />

solve the drainage problem, then the tubing should be replaced.<br />

To replace the drain pump tubing, perform the following steps:<br />

1 Turn the AC power switch to off (0) and disconnect the AC<br />

power cord.<br />

2 Remove the 1/16 inch O.D. PFA Teflon® tubing (3 lines) from the<br />

top inlet side of the drain pump. Then remove the 3 longer lines<br />

of FEP-lined Tygon® tubing from the outlet side of the pump.<br />

3 Release the tension screws for all three channels of the drain<br />

pump.<br />

4 Carefully pull out the connector fittings from the drain pump<br />

tubing holder.<br />

5 Remove the tubing from the fittings and replace with new<br />

tubing.<br />

6 Reattach the drain tubing with the connector fittings to the<br />

drain pump tubing holder (inlet and outlet).<br />

7 Reconnect the 1/16 inch O.D. PFA Teflon® tubing (3 lines) to the<br />

top inlet side, and the longer tubing to the outlet side (3 lines) of<br />

the drain pump.<br />

8 Secure the tension screws for all three channels of the drain<br />

pump.<br />

9 Adjust the tension screws as needed for a steady flow of drain<br />

liquid.<br />

6-9