Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Marin-5 Operator's Manual - CETAC Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Troubleshooting the <strong>Marin</strong>-5<br />
background ratios and elemental oxide levels (such as CeO with<br />
ICP-MS).<br />
• Liquid accumulation in spray chamber. Replace the peristaltic<br />
pump drain tubing and adjust the pump shoe pressure to be<br />
slightly firm. Inspect the pump tubing on a daily basis and replace<br />
as required especially when it becomes worn and flattened.<br />
3 Poor precision.<br />
• Poor precision is often due to improperly set nebulizer gas flow. If<br />
the imprecision is characterized by large, periodic drops in signal<br />
intensity, the cause may be an excessive sample flow rate. More<br />
concentrated samples (>1% dissolved solids) may also cause poor<br />
precision. In addition, such samples may require longer rinse-out<br />
times. Finally, readjustment or replacement of the peristaltic pump<br />
drain tubing may be necessary to improve precision.<br />
4 Nebulizer self-aspiration stops.<br />
• Make certain that the nebulizer gas is turned on, the nebulizer gas<br />
line is leak-free and the proper amount of gas pressure (at least 40<br />
psi) is being delivered to the nebulizer.<br />
• The nebulizer capillary may be irreversibly clogged or damaged.<br />
Contact <strong>CETAC</strong> <strong>Technologies</strong> for a replacement capillary.<br />
Plasma Problems<br />
Plasma is unstable or will not start.<br />
• The most likely cause of plasma instability is a gas leak in the<br />
<strong>Marin</strong>-5. Check all gas line connections. Make sure all connections<br />
are tight and correct. Very high gas flow may also cause plasma<br />
instability. Refer to Chapter 4, “Verifying Installation.”<br />
7-5