<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />

Using the <strong>Marin</strong>-5<br />

4 Turn off the power switch on the <strong>Marin</strong>-5.<br />

Temperature Controller Operation<br />

The temperature controllers for the spray chamber/glassware and the<br />

condenser system display the actual operating temperatures of the two<br />

components. The setpoint for each temperature controller can be viewed<br />

by simply pressing the blue-colored button labeled SET on the<br />

respective temperature controller. When the button is released, the<br />

actual temperature is again displayed. The setpoint temperature can be<br />

changed by following the steps below:<br />

1 Press and hold the SET button and press the up or down arrow<br />

until the desired setpoint is reached.<br />

2 Release the SET button and the actual temperature will be<br />

displayed.<br />

3 The temperature controller will now adjust the temperature to<br />

the new setpoint.<br />

The temperature range for the spray chamber is 50°C to 150°C,<br />

with a factory setting of 120°C. (75°C for ICP-MS)<br />

The temperature range for the condenser is -20°C to 10°C, with a<br />

factory setting of 3°C.<br />

Note:<br />

The temperature controllers are designed to allow adjustment over a<br />

certain range. You must contact <strong>CETAC</strong> Customer Support if you<br />

desire an operating temperature outside the specified range.<br />

Unauthorized changes in the temperature settings will immediately<br />

terminate the warranty.<br />

5–5