<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Using the <strong>Marin</strong>-5<br />
4 Turn off the power switch on the <strong>Marin</strong>-5.<br />
Temperature Controller Operation<br />
The temperature controllers for the spray chamber/glassware and the<br />
condenser system display the actual operating temperatures of the two<br />
components. The setpoint for each temperature controller can be viewed<br />
by simply pressing the blue-colored button labeled SET on the<br />
respective temperature controller. When the button is released, the<br />
actual temperature is again displayed. The setpoint temperature can be<br />
changed by following the steps below:<br />
1 Press and hold the SET button and press the up or down arrow<br />
until the desired setpoint is reached.<br />
2 Release the SET button and the actual temperature will be<br />
displayed.<br />
3 The temperature controller will now adjust the temperature to<br />
the new setpoint.<br />
The temperature range for the spray chamber is 50°C to 150°C,<br />
with a factory setting of 120°C. (75°C for ICP-MS)<br />
The temperature range for the condenser is -20°C to 10°C, with a<br />
factory setting of 3°C.<br />
Note:<br />
The temperature controllers are designed to allow adjustment over a<br />
certain range. You must contact <strong>CETAC</strong> Customer Support if you<br />
desire an operating temperature outside the specified range.<br />
Unauthorized changes in the temperature settings will immediately<br />
terminate the warranty.<br />
5–5