<strong>Marin</strong>-5 Operator’s <strong>Manual</strong><br />
Maintaining the <strong>Marin</strong>-5<br />
__________________________________________________________________<br />
Maintaining the Concentric Nebulizer<br />
The <strong>Marin</strong>-5 concentric nebulizer should not be removed from the unit<br />
while the plasma is on. Air intake through the nebulizer port can cause<br />
the plasma to extinguish.<br />
If the aspiration rate is extremely slow or if the nebulizer does not selfaspirate,<br />
the uptake capillary may be obstructed. To remedy this<br />
condition, replace the uptake line.<br />
To replace the uptake line, turn off power to the <strong>Marin</strong>-5 and disconnect<br />
the power cord. Allow the <strong>Marin</strong>-5 to cool for 30 minutes. Then open the<br />
front door and remove the nebulizer uptake line and replace with a new<br />
line. If a new uptake line does not solve the uptake problem, then there<br />
may be an obstruction in the nebulizer itself. The nebulizer may need to<br />
be replaced.<br />
After a sequence of sample analyses the nebulizer should be rinsed to<br />
prevent solid buildup on the nebulizer tip. A solution of 2% HNO3 may<br />
be aspirated for 5 minutes followed by deionized water for 5 minutes.<br />
__________________________________________________________________<br />
Main Fuse Replacement<br />
The main fuses are located in the AC power module fuse drawer located<br />
at the left rear of the <strong>Marin</strong>-5. To replace blown fuses:<br />
1 Turn the AC power switch to off (0) and disconnect the AC<br />
power cord.<br />
2 Use a small flat blade screwdriver to unlatch the fuse holder.<br />
3 Replace the defective fuse with a 5A, 250V 20mm long fuse.<br />
WARNING<br />
Use of a different fuse other than those specified can damage the<br />
electrical components of the <strong>Marin</strong>-5, constitute a fire hazard or result<br />
in personal injury.<br />