Troubleshooting the <strong>Marin</strong>-5<br />
3 Temperature controller stops illuminating.<br />
• Thermal safety switch tripped due to excessive temperature. Heat<br />
controller or solid-state relay is defective. Contact <strong>CETAC</strong><br />
Customer Service.<br />
Solution Drainage Problems<br />
1 Liquid waste is not removed from the <strong>Marin</strong>-5.<br />
• Adjust tension on drain pump to improve liquid waste flow.<br />
• Check for wear or damage to drain pump tubing. Replace tubing if<br />
necessary.<br />
2 Drain pump will not operate.<br />
• Contact <strong>CETAC</strong> Customer Service.<br />
Nebulizer / Plasma Problems<br />
1 Poor and/or unstable aerosol generation.<br />
• Insufficient nebulizer gas flow or pressure. The Argon<br />
nebulizer gas flow should be in a 0.5 - 1.1 L/min range. If such a<br />
flow range cannot be achieved, there may be insufficient nebulizer<br />
gas pressure. Increase inlet pressure (not to exceed 100 psi or<br />
700kPa.). Normal operating pressure for the <strong>Marin</strong>-5 is 40 psi<br />
argon pressure. If the ICP-AES or ICP-MS cannot achieve 40 psi,<br />
alternative plumbing may be required.<br />
• Uptake line may be obstructed or damaged. Replace the<br />
nebulizer uptake line.<br />
• Leak in the nebulizer gas line. Inspect the nebulizer gas line to<br />
make certain that all connections are correct and leak-free. If the<br />
line is leaking, repair with appropriate tubing.<br />
7-3