Troubleshooting the <strong>Marin</strong>-5<br />

3 Temperature controller stops illuminating.<br />

• Thermal safety switch tripped due to excessive temperature. Heat<br />

controller or solid-state relay is defective. Contact <strong>CETAC</strong><br />

Customer Service.<br />

Solution Drainage Problems<br />

1 Liquid waste is not removed from the <strong>Marin</strong>-5.<br />

• Adjust tension on drain pump to improve liquid waste flow.<br />

• Check for wear or damage to drain pump tubing. Replace tubing if<br />

necessary.<br />

2 Drain pump will not operate.<br />

• Contact <strong>CETAC</strong> Customer Service.<br />

Nebulizer / Plasma Problems<br />

1 Poor and/or unstable aerosol generation.<br />

• Insufficient nebulizer gas flow or pressure. The Argon<br />

nebulizer gas flow should be in a 0.5 - 1.1 L/min range. If such a<br />

flow range cannot be achieved, there may be insufficient nebulizer<br />

gas pressure. Increase inlet pressure (not to exceed 100 psi or<br />

700kPa.). Normal operating pressure for the <strong>Marin</strong>-5 is 40 psi<br />

argon pressure. If the ICP-AES or ICP-MS cannot achieve 40 psi,<br />

alternative plumbing may be required.<br />

• Uptake line may be obstructed or damaged. Replace the<br />

nebulizer uptake line.<br />

• Leak in the nebulizer gas line. Inspect the nebulizer gas line to<br />

make certain that all connections are correct and leak-free. If the<br />

line is leaking, repair with appropriate tubing.<br />

7-3