52DIGITAL-PANEL-MOUNTED DISPLAYMODULES for all applications• 2 temperature modules (covering temperature rangesfrom -50 up to +1150° C)• 3 pressure modules for barometer, vacuum meter, manometerfor absolute pressure, over/under pressure and pressuredifference measurements. Pressure range up to 10 bar• one voltmeter module with 3 integrated voltage rangesCommon specification for all modules:Display: 3½-digit LCD display, 13mm high (±1999 digit), scan rate: 3meas. per second, operating temperature: 0 to 50°C, atmospherichumidity: 0 to 85%r.h. (non-condensing), storage temperature: -10to +70°C, current supply: 9 - 12 V DC, dimensions: 38 x 76 x 22 mm(H x W x D), panel-cutout: 36 +0.5 x 73.2 +0.5 mm (H x W), panelthickness: max. up to 9.5mm. snap-on frame protruding only 1mm overfront plate - professional design, 3mm thick anti-reflex screenTEMPERATUREGPT 180TEMPERATURE MODULE for semiconductor sensor KTY 83-110Range: -50.0 up to +175.0° C / Resolution: 0.1° CAccuracy: approx. 1% f.s. / Power consumption: approx. 1 mASuitable sensors KTY 83-110: please refer to pages 89 and 93GPT 1155TEMPERATURE MODULE for thermocouple NiCr-Ni (type K)Range: -50 up to +1150° C / Resolution: 1° CAccuracy: better than 1 % from -20 up to +550 and from 920 up to 1150° C,550 up to 920 better than 1.5%Power consumption: approx. 0.35 mASuitable sensors type NiCr-Ni (type K) p.r.t. pages 14,15 and 90,91GTU 300/152 wire sensor with soldering pin plugPressureGPD 15 ABSDIGITAL BAROMETER / VACUUM METER MODULE (sensor not included)Range: 0 to 1100 mbar (hPa) absolute / Resolution: 1 mbarAccuracy module: 1 mbar ±1 digitAccuracy sensor: (sensor not included in scope of supply):±0.2% (typical) for linearity and hysteresis, ±0.4% for temperature driftfrom 0 to 50° C (typ. values for sensors compensated to module)Power consumption (incl. sensor) approx. 3.5 mASuitable sensors: (please order separately)SCX 15 ANC (pressure sensor)Upcharge for double accuracyGPD 30 RELDIGITAL MANOMETER for over/under pressure and pressuredifference (sensor not included)Meas. range: -1000 to +1999 mbar relative (referring to ambient pressure)Resolution 1 mbar / Accuracy module 1 mbar ±1 digitAccuracy sensor and power consumption as aboveSuitable sensors: (please order separately)SCX 30 DNC (pressure sensor)Upcharge for double accuracyGPD 150 RELDIGITAL MANOMETER for over/under pressure and pressuredifference (sensor not included)Range: -1.00 up to 10.00 bar relative (referring to ambient pressure)Resolution 0.01 bar Accuracy module 1 mbar ±1 digitAccuracy sensor and power consumption as aboveSuitable sensors: (please order separately)SCX 150 DNC (pressure sensor)Upcharge for double accuracyVOLTAGEGPV 220DIGITAL VOLTMETER, 3 integrated voltage ranges - others can be realisedby means of an external voltage divider (eg for mains voltage 230 V etc.)Ranges: ±199.9 mV, ±1999 mV, ±19.99 V integrated; (±199.9V or 1999Vcan be realised by means of an external voltage divider)Decimal point: any place selectableResolution: up to 100µV / Input impedance: 100MÐ resp. 1MÐAccuracy: 0.1% ±1 digit / T.C. value: 100 ppm/KPower consumption: approx. 100µA only (approx. 3000 hours withnormal 9V-battery)DIGITAL DISPLAY for all measuringtransducers 4 to 20 mA 2-wire, noauxiliary power requiredDigital panel module<strong>without</strong> auxiliary energy• for use in 4 to 20 mAoutput circuits ofmeasuring transducers• WITHOUT EXTERNALAUXILIARY SUPPLY• high operatingreliability• Cost reduction aspower supplies andtheir cables are nolonger requiredGTA 0420 (standard range)Large, high-contrast 3 1/2 digit LCD, 12.7 mm high; to either directly displayloop current or convert it into any desired value such as temperature,pressure, fill level, humidity, travel, weight, height, liquid flow, ppm, mg/l, %sat., etc..Snap-on, industrial panel-mounting type, anti-reflex screen 3 mm thick (notto be compared with unprotected glass covered display as used withcheap modules!)Minimum size: 38 x 76 x 22 mm (H x W x D). Devices can be stack-mountedat a distance of 38 mm.Standard printings available, eg. °C, %, V, mbar, bar, otherwise neutral.Specification:Input signal: 4 .. 20 mA, 2-wireDisplay ranges: 0,0 ... 100,0; 0,0 ... 199,9; -50,0 ... +50,0 (standard);any display range desired against upcharge (p.r.t. options)Decimal point: any place (soldering jumper)ine tuning: starting point at 4 mA and end point at 20 mA can each beshifted by ±50 digitsDisplay: 3½ digit LCD with ±1999 digits, 13 mm highScanning rate: 3 measurements per secondAccuracy: ±0.1% ±1digitTemperatue coefficient: 100 ppm / KOperating temperature: 0 to 50°CAtmospheric humidity: 0 to 85%r.h. (non-condensing)Storage temperature: -10 to +70°CDimensions: 38 x 76 x 22 mm (H x W x D)Panel cutout: 36 +0.5 x 73.2 +0.5 mm (H x W)Panel thickness: max. up to 9.5mm.Options:Any measuring range desired (against upcharge)(no upcharge for orders as of 10 pieces of the same range)urther displays <strong>without</strong> auxiliary supply:freely scaleable display for control panel mounting (24 x 48 mm)GIA0420 - p.r.t. page 50freely scaleable plug on displayGIA0420VO - p.r.t. page 51freely scaleable display for surface or wall mountingGIA0420WK - p.r.t. page 51
The new program controller - the temperature-set-point value shaping upExtremly comfortable operation is the main advantage of theprogram controller grado 964: as a standard all programparameters and sections can be adjusted via an integratedinterface from your PC.grado 964Specification:Inputs: universal input for- Pt100 (3-wire)- Thermocouples: J, K, R, S, T, B, L, N- DC linear mA: 0-20mA, 4-20mA- DC mV: 0-50, 10-50mV, 0-5, 1-5, 0-10, 2-10V(Please specify type desired on your order)Meas. range: Pt100: -200...800°C;Thermocouples type K: -200...1373°C,Type J: 0...761°C, type R and S: 0...1649°C,Type T: -200...262°C, type L: 0...762°CType B: 100...1824°C, type N: 0...1399°CResolution: temperature: 0.1, 1°C or 0.1, 1°Standard signals: scale freely adjustable -1999...9999.Accuracy: ± 0.25 % or ±1digitMeas. rate: 4 measurements / sec.Display: two lines, 4-digit LED-display,nominal value: 10mm high, red, set-point value:8mm, green.Outputs: (type: switching)Output 1: relay output(changeover, switching power: 2A, 240VAC)or logic output (please specify on your order)l DIN-format 48 x 48 mm program controllerl 4 programs, each program with up to 16 segmentsl PC-configuratorl or temperature and time controll RaPID uzzy Logictime controlled event contact for output 2 or 3l Program repeat can be set from 1 to 9999 and endlessl Timer for program start delay up to a max. of99 h 59 min.l Time basis for each program to be set inhours (up to 99 h 59 min., each segment)minutes (up to 99 min. 59 sec., each segment)l External START / HOLD functionl Separate locking code (for program and controlparametersOutput 2, 3 [optional]: relay output, logicoutput or linear (4-20mA, 0-20mA, 0-5V or0-10V) (please specify type desired on your order)Controller state: 2-level- or 3-levelcontrollerOutputs: (type: continuous)Output 1: linear output (4-20mA, 0-20mA,0-5V oder 0-10V(please specify type desired on your order)Output 2, 3 [optional]: relay-output, logicoutput or linear (4-20mA, 0-20mA, 0-5V or 0-10V)Controller state: continuous controlControl parameter: manual (P, PI, PD orPID-controller) or automatic by means ofpreliminary adjustment or indirect acting.Alarm types: process alarm (high undlow), alarm in case of deviations, bandloss alarmInterface [option]: serial, RS485,Baud rate: 9600, 4800, 2400, 1200Housing: 48 x 48 x 110 mm (W x H x D)Panel cutout: 45 x 45 mmMounting: by means of clamping frameIP rating: front side IP65, connections IP20Electric connection: screw-type terminalsOperating temperature: 0 ... +55 °CStorage temperature: -20 ... +80 °CNoise immunity (EMC): conforming toEN50<strong>08</strong>2-2 and EN50<strong>08</strong>1-2Voltage supply:90...264 VAC 50/60Hz.Option: 20...50VAC; 22...65VDC (upcharge)Power input: approx. 4 WattOptions:Cont. control responseadd. output (relay)add. output (logic)add. output (linear)RS485-interfacePower supply: 22-50VAC/DCStart-/Hold-functionPC-software:Configuration- and program-software incl.RS232-adapter cablePC-software:The PC software is installed underWindows. A special adapter cable -RS232/485 is used to connect the PC andprogram controller. By clicking on thepreset parameters these values will betaken over into the program. Parametersmay also be entered directly via thekeyboard. Switching over from theprogram level to the configuration level ispossible. You may also open bothwindows to appear simultaneously onyour screen and thus be able to constantlymonitor the configuration of the controller.Configuration-softwareProgram-software53